Brief description: Clean laminar flow hood is a new type of laminar flow hood designed according to the actual situation of clean rooms, suitable for industries such as electronics, precision instruments, medical food, etc. Can be lifted or supported for use. $r $n 1. Product Features $r $n1. The airflow form adopts the vertical unidirectional flow purification principle, which is composed of a centrifugal fan and a static pressure box
Clean laminar flow hood1、 Product Features
1. The airflow form adopts the principle of vertical unidirectional flow purification, consisting of a centrifugal fan, a static pressure box,A combination structure consisting of a liquid tank high-efficiency filter and a uniform flow membrane.
2. Purification level: Level 100 (Federal Standard 209E)
3. Average wind speed: 0.25-0.45m/s
4. Power supply voltage: 220V, 50HZ
5. Noise: ≤ 62dB (A)
Clean laminar flow hood2、 Parameter configuration
|
model
parameter
|
FXCL-700 |
FXCL-1300 |
FXCL-1900 |
FXCL-2500 |
Purification area (mm) |
650*610 |
1250*950 |
1850*1220 |
2450*1220 |
Dimensions (mm) |
700*750*750 |
1300*1100*750 |
1900*1400*750 |
2500*1400*750 |
Power consumption (KW) |
0.4 |
0.8 |
1.2 |
1.6 |
Size and quantity of liquid tank filters |
700*500*90mm*1 |
1000*500*90mm*2 |
1000*500*90mm*3 |
1000*550*90mm*4 |
lighting |
300LUX |
300LUX |
300LUX |
300LUX |
DOP detection port |
contain |
contain |
contain |
contain |
differential pressure gauge |
contain |
contain |
contain |
contain |
3、 Structural features