Chromatographic Column Mo'er Chemical CC553-C18-AQ

NegotiableUpdate on 05/10
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

Caprisil C18-AQ is an aqueous column resistant to pure aqueous phases, suitable for the analysis of high water flow mobile phases and providing good selectivity for the analysis of highly polar compounds. There are hydrophilic groups on the surface of the chromatographic packing, which increase the interaction. When separating polar compounds in high water phase systems, there will be no change in retention characteristics, which greatly improves the theoretical plate number and chromatographic peak separation performance. The chromatographic column CC553-C18-AQ is suitable for the separation of hydrophilic compounds; $r $n ● Pure water solution can be used as the mobile phase; $r $n ● with conventional ODS chromatography

Product Details

MORHCHEM is a professional manufacturer of consumables related to chromatography columns. The company's headquarters is located in California, USA. After more than 30 years of continuous innovation and challenges, it has developed Caprisil,NexHgel,ItolSep,Guard column,Synersorb,Spherigel,HyperStar,Prep,CapriChiarl A series of HPLC chromatography columns.

MORHCHEM is committed to helping customers solve various challenges encountered in the field of analytical chemistry, from routine testing to complex research and development projects. Our clients include pharmaceutical and biotech companies, clinical diagnostic laboratories, universities, research institutes, and government agencies.


Chromatographic Column Mo'er Chemical CC553-C18-AQ


Caprisil C18-AQ is an aqueous column resistant to pure aqueous phases, suitable for the analysis of high water flow mobile phases and providing good selectivity for the analysis of highly polar compounds. There are hydrophilic groups on the surface of the chromatographic packing, which increase the interaction. When separating polar compounds in high water phase systems, there will be no change in retention characteristics, which greatly improves the theoretical plate number and chromatographic peak separation performance.

Suitable for the separation of hydrophilic compounds;

The mobile phase can use pure water solution;

Has selectivity different from conventional ODS chromatography columns.

色谱柱漠迩化学CC553-C18-AQ


Chromatographic Column Mo'er Chemical CC553-C18-AQ


Chromatography column: CE553-C18-AQ

Flow rate: 1ml/min

Injection volume: 10ul

Mobile phase: A: 0.1%H?PO?B: Acetonitrile

Time A B

0-10 98 2

20 90 10

30 80 20

70 45 55

75 20 80

83 20 80

Detector: 221nm


Sample spectrum:

色谱柱漠迩化学CC553-C18-AQ


Instructions for using reverse phase chromatography column


1、 Conventional flushing method under mobile phase conditions

1. Before use: If methanol water or acetonitrile water is used as the mobile phase, the sample can be directly equilibrated with the mobile phase for 40 minutes before baseline stabilization.

2. After use: Rinse with pure methanol or acetonitrile water (70:30) for more than 40 minutes and store.

2、 The mobile phase contains buffer salts

1. Transition before use: Transition with 90% water and organic phase for 40-60 minutes. Then, after 40 minutes of equilibration under mobile phase conditions, the sample is taken (if under gradient conditions, the initial gradient equilibrium is followed).

2. Rinse after use: Rinse with 90% water and organic phase for at least 60 minutes. Then rinse with pure methanol or acetonitrile water (70:30) for more than 40 minutes and store.

3、 The mobile phase contains ion pair reagents

1. Transition before use: Transition with 90% water and organic phase for 40-60 minutes. Then equilibrate the sample under mobile phase conditions for 120 minutes (if it is under gradient conditions, follow the initial gradient equilibrium).

2. Rinse after use: Rinse with 90% water and organic phase for at least 180 minutes. Then rinse with pure methanol or acetonitrile water (70:30) for more than 40 minutes and store.

Note:

1.Factory storage conditions:7030=Acetonitrile: Water.

2.It is recommended to use a dedicated column for chromatography columns containing ion pair reagents.

3.It is recommended to store it in the factory solvent condition for long-term non use, and it is strictly prohibited to store it in a mobile phase containing salt.