Welcome Customer !

Membership

Help

Shanghai Gelan Instrument Equipment Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shanghai Gelan Instrument Equipment Co., Ltd

  • E-mail

  • Phone

  • Address

    5th Floor, Building 5, Phase II, Shanghai University Science and Technology Park, 599 Huiwang East Road, Jiading District, Shanghai

Contact Now

Chip online total iron/iron ion water quality analyzer

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The PROCON5200 chip online total iron/iron ion water quality analyzer can measure and control the concentration of nickel ions in water. It uses spectrophotometry to detect the total dissolved nickel content in water. The analyzer consists of a control unit and a measurement and analysis unit that integrates measurement chambers, valves, metering pumps, and pipelines. The host microprocessor can fully control the measurement process, from sample injection, flushing, to reagent pumping, and then rely on the photoelectric system to complete the detection, providing support for monitoring the concentration of nickel ions in chemical wastewater.

Product Details

Chip online total iron/iron ion water quality analyzerPROCON5200

Chip online total iron/iron ion water quality analyzerPROCON5200Capable of measuring and controlling the concentration of nickel ions in water. It uses spectrophotometry to detect the total dissolved nickel content in water. The analyzer consists of a control unit and a measurement and analysis unit that integrates measurement chambers, valves, metering pumps, and pipelines. The host microprocessor can fully control the measurement process, from sample injection, flushing, to reagent pumping, and then rely on the photoelectric system to complete the detection, providing support for monitoring the concentration of nickel ions in chemical wastewater.

Product Features:

● Automatic measurement, including self check and drift compensation;

● Different measurement ranges to choose from;

● The touch screen is easy to operate;

The limit and alarm values can be adjusted;

● Programmable 0/4-20mA analog output;

● USB data collection (optional);

● Adjustable analysis interval;

● External start stop analysis is possible;

No additional calibration is required;

External plug connection (IP65) is used for alarm/limit/analysis relays, external start stop, and analog output.

Technical parameters:

type number

PROCON5200

measurementmethod

Total nickel

measurementscope

0.01-1.00ppm

resolution

0.01ppm

precision

3% final value

repetitiveness

2% final value

Zero stability

auto-adjust

size

640×315×190mm(H×W×D)

display

240×128, touch screen

output

0/4-20mA, Maximum load of 500 Ω

Analysis Cycle

4-7 minutes with rinsing

interval1-99min,1-99min

temperature

5-45℃

power supply voltage

110-230V(50/60Hz)

electrical connection

1.5mm2 circular plug connection

relay

1x alarm, zero potential contact 230v/50Hz, 3A

1x limit, zero potential contact 230v/50Hz, 3A

1x analysis state, zero potential contact 230v/50Hz, 3A

remote control

Zero potential contact,18V DC,ca.4 mA

Protection level

IP65

energy consumption

approximately16VA