| instrument parameters | Parameter Description | measuring method | notes | |
| measureQuantity range | 10-1000.0 mg/L | National standard GB11914-89 "Water quality - Determination of chemical oxygen demand - Potassium dichromate" | Automatic alarm for instrument abnormalities; Automatic saving of power-off data; 7-inch 60000 color TFT touch screen display and command input; After abnormal reset and power failure, the instrument will automatically discharge residual reactants and restore working status. | |
| Detection limit | 0.001mg/L | |||
| refinedAccuracy | When ≥ 100mg/L, not exceeding ± 15%; When<100mg/L, not exceeding ± 5mg/L | |||
| reproducibility | ±5% | |||
| measureQuantity cycle | The measurement cycle is 35 minutes | |||
| pickSample cycle | Time interval (1-65530 minutes can be set arbitrarily) and 24-hour hourly measurement mode | |||
| schoolQuasi cycle | Time interval (1-65530 minutes can be set arbitrarily) and 24-hour hourly measurement mode | |||
| Maintenance Cycle | monthly | |||
| letterNumber output | Standard RS-232 (RS485 optional) and 4-20mA (0-5V output optional) | |||
| ringEnvironmental requirements | For indoor environments with adjustable temperature, it is recommended to have a temperature range of+5-35 ℃; Humidity ≤ 90% (non condensing) | |||
| electricitysource | AC220×(1±0.1)V,50×(1±0.05)Hz,10A | |||
| chicun | 600mm in length, 500 in width, and 1500 in height |
|||