The wavelength calibration standard substance for the Starna HPLC detector in the UK is used for wavelength calibration of HPLC diode array detectors that can be placed in a 10mm optical path colorimetric cell at the sample detection position; A 10% v/v perchloric acid solution of erbium oxide (1% m/v) is sealed in a 10mm optical path far ultraviolet quartz colorimetric cell that is fused together. All Starna reference materials are traceable to NIST and have calibration certificates accredited by UKAS ISO/IEC 17025 and ISO Guide 34
Traceability:allStarnaThe standard substances are allNISTTraceable and withUKAS ISO/IEC 17025andISO Guide 34Recognized calibration certificate
Purpose: This product can be placed in the sample testing position10mmThe optical path colorimetric cellHPLCWavelength calibration of diode array detector;
Product description:Erbium oxide(1% m/vof10%v/vHigh chloric acid solution, sealed in a molten one-piece molding10mmIn the far ultraviolet quartz colorimetric cell.
Calibrate with a perchloric acid solution of erbium oxideHPLCThe wavelength of a diode array detector is a method specified in many documents. High chloric acid solution of erbium oxide at wavelength230nm—260nmWithin the spectral range, there are9A distinct characteristic peak.

9The wavelength of each peak is approximately:255, 364, 379, 405, 449, 487, 522, 541, 652.
typicalHPLCVerification method (e.g:WatersWaters)Chinese usage255, 379and522 nmVerify the peaks at three wavelengths
Explanation: The above certified wavelength values are for reference only, and the actual values are determined by the spectral bandwidth. The certification certificate provided with the product specifies1.00and5.00 nmAuthentication wavelength under bandwidth, if required by the customer,StarnaCan provide authentication values for other bandwidths and wavelengths.
Product Code
Reference material RM-ER for perchloric acid solution of erbium oxide