Welcome Customer !

Membership

Help

Shanghai Aiyan Biotechnology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Shanghai Aiyan Biotechnology Co., Ltd

  • E-mail

  • Phone

  • Address

    No. 28, Lane 412, Jinxiu East Road, Pudong New Area, Shanghai

Contact Now

Bio rad dual 48 well gradient PCR instrument C1000

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The bio rad Bole dual 48 well gradient PCR instrument C1000 has replaceable reaction modules, including 96x0.2ml gradient single slot module, 2x48x0.2ml dual slot gradient module, 384 well single slot gradient module, quantitative PCR gradient module, etc.

Product Details

Performance description of bio rad dual 48 well gradient PCR instrument C1000

Equipped with replaceable reaction modules, including 96x0.2ml gradient single slot module, 2x48x0.2ml dual slot gradient module, 384 well single slot gradient module, quantitative PCR gradient module, etc

8.5 High resolution ultra large color LCD display screen, with options for graphic programming, text programming, or automatic programming

Dual slot temperature gradient module, capable of independently running two different programs

Equipped with dynamic temperature gradient function, it can operate at 8 different temperatures simultaneously, with the same incubation time for each temperature

Semiconductor heating and cooling method

Bio rad dual 48 well gradient PCR instrument C1000 is mainly used for in vitro nucleic acid fragment amplification with dynamic temperature gradient function

Technical parameters of Bole C1000PCR instrument:

96 blue-green LED sequentially excited samples greatly reduced fluorescence crossover

Suitable for 96 × 0.2ml reaction tubes or 12 0.2ml 8-linked reaction tubes.

It can support multiplex PCR, which not only saves reagents and time, but also makes it possible to use internal controls, thereby further controlling the efficiency of sample PCR.

perfectTemperature control performance and gradient PCR function.

OpticonMonitor analysis software can easily set up reactions such as RT-PCR and multiplex PCR, as well as perform functional analyses such as DNA quantification, gene expression, and allele analysis.

Accurate and reliable PCR instrument:

Opticon's C100PCR instrument is based on the mature Bio Rad product PTC-200, which features semiconductor refrigeration technology;

Three temperature sensors, dual zone temperature control;

Bole C1000PCR instrument

Accurate temperature control: The accuracy is ± 0.3 ℃ within 30 seconds at 90 ℃, even considering the influence of PCR reaction solution volume on temperature changes;

Good uniformity: Within 30 seconds at 90 ℃, the uniformity is ± 0.4 ℃;

96 reaction sample sizes;

Fast temperature rise and fall speed: up to 3.0 degrees per second;

There is a calculated temperature control mode: during the temperature rise and fall process, it does not exceed the head, thus ensuring the repeatability of the experiment;

Temperature gradient performance: The sample module can have 12 different temperatures simultaneously from left to right, forming a temperature gradient. The gradient temperature difference is set to a minimum of 1 ℃ and a maximum of 24 ℃. Through one experiment, the annealing temperature or denaturation temperature can be determined, which not only saves time but also reduces costs.

Fluorescence detection system combining dynamic and static detection:

Four independent excitation channels are excited using light sources of different wavelengths (light-emitting diodes, LEDs) to achieve the highest excitation efficiency for each channel;

Four independent detection channels use detectors with different photosensitive ranges (photodiodes, PDTs) for detection, achieving the highest emission detection efficiency for each channel;

The light source does not generate heat, has a long service life, and does not need to be replaced for life;

The position of the light source and detector is relatively fixed, ensuring that the optical path * is consistent and * between the holes

Interference issues;

The quantitative PCR instrument with the shortest optical path, with a total optical path of less than 8cm and the least light propagation loss, detects over 10 fluorescent dyes including FAM, SYBRGreenI, VIC, HEX, Joe, TAMRA, Cy3, Cy5, TexasRed, Rox, etc.

Wavelength range: 450-730nm;

Wide linear range: up to 10 orders of magnitude, capable of detecting single copy templates;

The resolution of Jijia can distinguish between 32, 64, 128, 256, and 1024 copies;

technical parameters

C1000
display screen 12 x 9 cm, High resolution, full-color
interface 5 USB-A
1 USB-B
Compatible with USB flash memory is
Compatible with USB mouse is
Programming Options Graphics, Text, Automatic Programmer
Automatic programmer is
report Can output running logs and system error logs
safety performance Protection mode, login options, folder protection
Temperature control mode Logic algorithm and module temperature control
Upgrade to a quantitative PCR instrument is
PCR certificate have
Online software Windows CE 6.0
Adapt to workstations is
store More than 1000 programs, USB expansion
heat preservation function have
Size (W X D X H) 13”x18”x8”
weight 23 lbs
Temperature gradient accuracy* ±0.2°C
Uniformity of temperature gradient rows* ±0.4°C?
Temperature gradient range* 30-100 ° C, not lower than room temperature of 30 ° C
Temperature gradient temperature difference range 1-24°C
voltage range 100-240VAC (voltage does not need to be adjusted within this range)
frequency 50–60 Hz
output power zui大 400 W
fuse 2, 6.3 A, 250 V, 5 x 20 mm

Note: Only 96 hole and dual 48 hole heating modules have temperature gradient function.

Technical parameters of the adapted heating module


96 hole quick module 48/48 hole quick module 384 holes
Sample capacity 96x 0.2mL 2x48x 0.2mL 1 384 well reaction plate
Temperature gradient function have have nothing
Temperature gradient range 30-100°C X
Temperature gradient temperature difference range 1-24°C X
Temperature gradient accuracy ±0.2°C X
Uniformity of temperature gradient ± 0.4 ° C (maintained for 10 seconds at 90 ° C) X
Hot cover temperature range 0-110°C
The maximum rate of temperature increase and decrease 5 ° C/second 4 ° C/second 2.5 ° C/second
temperature range 0-100°C
Temperature accuracy ±0.2°C
Temperature uniformity ± 0.4 ° C (reaching 90 ° C in 10 seconds)