BZDN-2Type semi micro steam nitrogen analyzer
summary:
As an important element in coal, nitrogen in coal generates nitrogen oxides during combustion, which seriously pollute the atmosphere and easily form acid rain, causing soil acidification. At the same time, nitrite is an important carcinogen. In power industry, the determination of nitrogen content in coal can be used to calculate the combustion heat of coal, theoretical combustion temperature, composition of combustion products and heat balance.
Usage conditions:
(1) Temperature: 0-40 ℃
(2) Relative humidity: ≤ 80%
(3) Power supply: Voltage: AC220 ± 22V
Frequency: 50 ± 0.5Hz
(4) Carrier gas: helium, purity higher than 99.8%
(5) Power: not less than 4.5KW
(6) No strong magnetic field interference
Basic parameters:
(1) Measurement range: 0.01-50%
(2) Sample size: about 100mg;
(3) Single measurement time: about 35 minutes;
(4) Rated power consumption: 4000W
Technical requirements:
(1) Repeatability (in accordance with GB/T 19227-2008) ≤ 0.08%
(2) Temperature control accuracy: set value ± 10 ℃
(3) Temperature control range: hydrolysis furnace: room temperature~1200 ℃
Distillation furnace: room temperature~500 ℃
Steam generator: room temperature~500 ℃
Standard configuration: BZDN-2 host