| Technical parameters of BSIC-160CH battery constant temperature test box integrated machine: | |||
| project | Required parameters | ||
| Specifications | volume | 200L | |
| Internal cavity size | 500mm*500mm*800mm | ||
| Overall dimensions | 700mm*1100mm*1720mm | ||
| weight | ≤200kg | ||
| performance | temperature range | 0℃~60℃ | |
| temperature fluctuation | ≤± 0.5 ℃ (unloaded) | ||
| temperature deviation | ± 1 ℃ (unloaded) | ||
| warm-up time | 25 ℃ to 60 ℃ ≤ 30min (no load, average nonlinearity) | ||
| cooling time | 25 ℃ to 0 ℃ ≤ 50min (no load, average nonlinearity) | ||
| structure feature |
Outer wall material | High quality cold-rolled steel plate, surface spray and paint treatment | |
| Inner wall material | Stainless steel plate SUS304 | ||
| Box insulation material | Insulation board+polyurethane foam (insulation thickness 60mm) | ||
| Box door | Central control anti fog tempered glass+frame | ||
| Lead hole | 4 pieces with a diameter of 80mm (located on both sides of the box) | ||
| caster | 4 (with brakes) | ||
| Battery Cell Tray | Insulated cell tray with 4 layers, load-bearing capacity: 10kg/layer | ||
| lighting | LED lighting fixtures | ||
| Control Panel | Touch Control Button | ||
| refrigeration system |
refrigeration compressor | Fully enclosed piston compressor | |
| cooling method | air-cooled | ||
| Control method | Frost free control ensures long-term operation | ||
| refrigerant | R134a | ||
| welding process | Nitrogen protection welding | ||
| Electrical Control system |
controller | Touch screen controller | |
| setting method | touch-sensitive | ||
| Control method | Forced circulation ventilation balanced temperature control method. The control system controls the output of the heater through PID self calculation based on the set temperature value, thereby achieving dynamic balance | ||
| communication method | 485 communication | ||
| Test box safety protection device | Leakage protection, short circuit protection, and abnormal operation of circulating fans, etc | ||
| cell Specifications And release Place the square formula |
Cell specifications | Button type battery 160CH | |
| Cell placement method | Four layer placement, each layer is 40CH | ||
| Battery Cell Tray | The battery cell tray is made of electrically insulated electric wood material | ||