BINDER M115 drying oven
Product Details

BINDER M115 drying oven

1、 Product Introduction:

1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃

2、APT.lineTMPreheating chamber technology

3. Adjustable fan speed

4. Programmable air valve

5. LCD color controller using interval programming

6. 2 chrome plated sockets

7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm

8. Computer interface: RS 422

2、 Technical parameters:

Product Model M115-230V1
Choose a model standard
Item code 9010-0202
measurement
Internal volume [L] 115
Net weight of equipment (unloaded) [kg] 89
Maximum total load [kg] 50
Maximum load per partition [kg] 20
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 835
Net weight at height [mm] 865
Depth net weight [mm] 645
Wall to side distance [mm] 100
Wall distance from rear [mm] 160
internal dimensions
Width [mm] 600
Height [mm] 480
Depth [mm] 410
Data related to the environment
Energy consumption at 150 ℃ [Wh/h] 544
firmware
Number of partitions (hours/maximum) 2/6
Temperature performance data
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] 300
Temperature deviation is 150 ℃ [± K] 1.5
Temperature fluctuation at 150 ℃ [± K] 0.3
Heating time to 150 ℃ [min] 16
The recovery time after opening the door in 30 seconds is 150 ℃ [min] 3
Ventilation data
Air exchange rate at 150 ℃ [x/h] 96
Electrical data
Rated voltage [V] 230
Power frequency [Hz] 50/60
Rated power [kW] 1.6
Phase (rated voltage) 1~

1·All technical data is only applicable to standard specifications of unloaded equipment under ambient temperature of 22 ± 3 ℃ and ± 10% power supply voltage fluctuation. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.