- Phone
-
Address
Room 1806, Fortune Plaza South, 5616 Cao'an Road, Shanghai#
Shanghai Huanao Technology Co., Ltd
Room 1806, Fortune Plaza South, 5616 Cao'an Road, Shanghai#
BINDER FP720 drying oven
1、 Product Introduction:
1. Temperature range: ambient temperature plus 5 ℃ to 300 ℃
2 APT.line ™ Preheating chamber technology
3. Adjustable fan speed
4. Adjustable exhaust valve
5. Controller using interval programming and real-time programming
6. 2 chrome plated sockets
7. Independent adjustable temperature safety device level 2 (DIN 12880) with optical alarm
8. Computer interface: RS 422
2、 Technical parameters:
| Product Model | FP720-400V1 | FP720UL-208V1 |
| Choose a model | standard | standard |
| Item code | 9010-0267 | 9010-0268 |
| measurement | ||
| Internal volume [L] | 720 | 720 |
| Net weight of equipment (unloaded) [kg] | 194 | 194 |
| Maximum total load [kg] | 120 | 120 |
| Maximum load per partition [kg] | 45 | 45 |
| The dimensions of the casing do not include additional components and connectors | ||
| Net weight width [mm] | 1235 | 1235 |
| Net weight at height [mm] | 1530 | 1530 |
| Depth net weight [mm] | 865 | 865 |
| Wall distance from rear [mm] | 160 | 160 |
| Wall to side distance [mm] | 100 | 100 |
| internal dimensions | ||
| Width [mm] | 1000 | 1000 |
| Height [mm] | 1200 | 1200 |
| Depth [mm] | 610 | 610 |
| Data related to the environment | ||
| Energy consumption at 150 ℃ [Wh/h] | 1320 | 1320 |
| firmware | ||
| Number of partitions (hours/maximum) | 2/16 | 2/16 |
| Temperature performance data | ||
| The temperature range exceeds the indoor temperature by 5 ℃ to [℃] | 300 | 300 |
| Temperature deviation is 150 ℃ [± K] | 2 | 2 |
| Temperature fluctuation at 150 ℃ [± K] | 0.3 | 0.3 |
| Heating time to 150 ℃ [min] | 39 | 39 |
| The recovery time after opening the door in 30 seconds is 150 ℃ [min] | 20 | 20 |
| Ventilation data | ||
| Air exchange rate at 150 ℃ [x/h] | 12 | 12 |
| Electrical data | ||
| Rated voltage [V] | 400 | 208 |
| Power frequency [Hz] | 50/60 | 60 |
| Rated power [kW] | 5 | 5 |
| Equipment safety device [A] | 3 x 16 | 3 x 16 |
| Phase (rated voltage) | 3~ | 3~ |
1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. The technical data assumes a fan speed of 100%. We reserve the right to make technical changes.