BINDER BD23 Biochemical Incubator

Product Details

BINDER BD23 Biochemical Incubator

1、 Product Introduction:

1. Temperature range: ambient temperature plus 5 ℃ to 100 ℃

2 APT.line ™ Preheating chamber technology

3. Free convection

4. Adjustable exhaust valve

5. Controller with time function

6. Interior door made of safety glass (ESG)

7. 2 chrome plated sockets

8. Stackable

9. Independent adjustable temperature safety device, Class 3.1 (DIN 12880), equipped with optical temperature alarm

10. Computer interface: RS 422

2、 Technical parameters:

Product Model BD023-230V1 BD023UL-120V1
Choose a model standard standard
Item code 9010-0187 9010-0189
measurement
Internal volume [L] 20 20
Net weight of equipment (unloaded) [kg] 27 27
Maximum total load [kg] 25 25
Maximum load per partition [kg] 12 12
The dimensions of the casing do not include additional components and connectors
Net weight width [mm] 435 435
Net weight at height [mm] 495 495
Depth net weight [mm] 520 520
Wall distance from rear [mm] 160 160
Wall to side distance [mm] 100 100
internal dimensions
Width [mm] 222 222
Height [mm] 330 330
Depth [mm] 277 277
number of doors
outer gate 1 1
Data related to the environment
Energy consumption at 37 ℃ [Wh/h] 11 11
firmware
Number of partitions (hours/maximum) 2/4 2/4
Temperature performance data
The temperature range exceeds the indoor temperature by 5 ℃ to [℃] 100 100
The temperature deviation is 37 ℃ [± K] 0.5 0.5
Temperature fluctuation at 37 ℃ [± K] 0.2 0.2
Heating time to 37 ℃ [min] 59 59
The recovery time after opening the door in 30 seconds is 37 ℃ [min] 20 20
Electrical data
Rated voltage [V] 230 120
Power frequency [Hz] 50/60 60
Rated power [kW] 0.2 0.2
Equipment safety device [A] 12.5 12.5
Phase (rated voltage) 1~ 1~

1. All technical data is only applicable to standard specifications of unloaded equipment at an ambient temperature of 22 ± 3 ℃ and a power supply voltage fluctuation of ± 10%. Determine temperature data according to BINDER factory standards and DIN 12880:2007, and adapt to the recommended wall gap of 10% of the internal chamber height, width, and depth. For series devices, all data are typical averages. We reserve the right to make technical changes.