BCY-2A Wireless Multi parameter Tester for Mining Water Pump
Overview
BCY-2AWireless multi parameter tester for mining water pumpProvided thermodynamic methods andHydraulic method is two methods for measuring pump efficiency, which do not require stopping the pump during measurement, only measuringThe inlet and outlet temperature difference (or flow rate), pressure difference, and electrical parameters of the measuring pump can be used to measure the efficiency, head, and flow rate of the pump. This instrument is easy to use and has high measurement accuracyHigh, suitable for testing, arbitration, and technical renovation of pump stations.
BCY-2AWireless multi parameter tester for mining water pumps, utilizing inspectionMeasurement, signal analysis, computer technology, etc. can scientifically detect and diagnose the working status and performance of water pumps, providing scientific basis for improving pump performance. Widely used in measuring the efficiency of water supply and drainage equipment pumps in coal mines, oil fields, non coal mines, and factories, it can measure the temperature difference between the inlet and outlet of the pump, inlet pressure, outlet pressure, head, pipe network efficiency, pump efficiency, unit consumption, and flow rate. It is an ideal portable intelligent measuring instrument for online measurement and evaluation of centrifugal pump performance.
1.1 Explosion proof form of equipment
The explosion-proof type is: mining intrinsic safety type.
1.2 Environmental conditions that can work normally
a)Environmental temperature:-40℃~60℃;
b)Humidity:≤98%;
c)Atmospheric pressure:80 KPa~116 KPa;
d)In situations where there is no significant vibration or impact;
e)Explosive hazardous areas containing gas and coal dust;
f)A place without corrosive gases that damage insulation.
1.3 Capable of withstanding storage and transportation environmental conditions
a)High temperature:+60℃;
b)Low temperature:-40℃;
c)Humidity:≤95%;
d)Vibration:20m/s2;
e)Impact:300 m/s2.
2 Main performance indicators
2.1 Key Metrics
|
project parameters |
Basic Parameters |
||
Fanenclose |
resolution |
mistakepoor |
|
inlet temperature(℃) |
0~100 |
0.01 |
±0.20 |
temperature difference(℃) |
0~5 |
0.001 |
±0.005 |
inlet pressure(MPa) |
-0.1~1.5 |
0.001 |
±0.002 |
outlet pressure(MPa) |
-0.1~20 |
0.001 |
±0.020 |
Pump efficiency(%) |
0~100 |
0.01 |
±0.5% |
unit consumption(kWh/t.100) |
0~10 |
0.001 |
±0.5%rdg |
head(m) |
0~2000 |
0.01 |
±0.2%rdg |
traffic(m3/h) |
0~9999.99 |
0.01 |
±1%rdg |
ambient temperature(℃) |
-40~120 |
0.01 |
±0.20℃ |
ambient humidity(%RH) |
0~100 |
0.1 |
±3.0%RH |
atmospheric pressure(kPa) |
0~130 |
0.01 |
±0.04kPa |
power(KW)(Optional) |
0~1500 |
0.001 |
±0.5%FS |
VoltageV) (Optional)) |
0~1000 |
0.001 |
±0.2% FS |
electric currentA) (Optional)) |
0~500 |
0.001 |
±0.2% FS |
l The host and all sensors have obtained the mining product safety mark certificate And explosion-proof certificate; The equipment composed of the host and sensors has passed the intrinsic safety inspection and obtained the mining product safety mark certificate;
l All sensors and hosts transmit data wirelessly No need to connect wired cables on site, effectively solving the disadvantages of complicated on-site connection and wiring. The operation is simple and convenient;
l The host is an intrinsically safe tablet computer that can be used for multiple purposes, including real-time data collection, processing, and moreDisplay on-site data, all data is directly saved to the host, and test reports that meet user needs can be issued. The software design is more user-friendly,The host is powered by lithium manganese oxide batteries, and can be used for up to normal testing time 8 aMore than hours;
l All sensors have been updated to integrated wireless sensors, which are lightweight and compact,lithium manganese oxideBattery powered, low self discharge rate for easy maintenance, and can be used for normal testing timelong 10 More than one hour;
l adopt LoRa TM Spread spectrum modulation technology and independently developed wireless communication groupNetwork technology,Effectively solved the shortcoming of high delay in synchronous transmission of multiple modules, adoptingThe wireless real-time transmission technology used has high anti-interference ability and low error rate, with a long transmission distance. In response to on-site interference problems, adaptive frequency hopping technology is adopted, which has high anti-interference and anti attenuation capabilities;
l Two methods, thermodynamic and hydraulic, are provided for testing water pumps. The optional ultrasonic flowmeter can directly measure the liquid inside the pipeline Flow rate and flow rate;
l It can measure and calculate the power consumption per ton of water per hundred meters of the water pump unit, achieving energy savings for the water pumpCapable of monitoring;
l Can draw the operating curve diagram of the drainage system and issue a test report;
l The instrument consists of a collection tablet, wireless environment collection sensor, and wireless temperature difference sensorComposed of modules and wireless oil pressure sensors.
1.1 Main Interface
Between on-site tests, the first step is to determine the measurement method to be used:
If using thermodynamic method, please arrange GPD1.5W Mining intrinsically safe wireless voltageForce sensorsGPD20W Mining intrinsically safe wireless pressure sensorGWB100WMining intrinsically safe wireless temperature sensorCDZ3 Motor electrical parameter tester(optional);
If using hydraulic methods, please connect GPD1.5W Mining intrinsically safe wireless pressure sensorGPA20W Mining intrinsically safe wireless pressure sensor, mining wirelessultrasonic flowmeter(Optional))TheCDZ3 Motor electrical parameter tester(Optional));
press KJD3.7-A The power button on the right side of the mining intrinsically safe tablet computer,After waiting for the instrument to start, click on the 'Water Pump Wireless Multi parameter Tester' to enter the testing processThe main interface of the test software is shown in the following figure:

1.1 sensor
Click the 【 Sensor 】 button in the main interface to open the sensor interface,
On this interface, sensors and tablets can be automatically paired for communication, forming pairsThe post power indicator light is green, and communication failure is gray, as shown in the following figure:

The meanings of the buttons in the interface are introduced as follows:
Current frequency: Set and display the wireless frequency signal currently used by the tablet computerRoad number; The optional range of frequency channels is 1、2、3If you want to choose to use the newThe frequency channel number needs to be deselected as' Continuous Reading Battery 'before it can be inputtedEnter a new frequency number. It is generally not recommended to manually modify the frequency here.
Select: Select or deselect a sensor, and the selected sensor will beTest the interface for use.
Status: Green indicates normal communication, gray indicates unused or lost communicationDefeated.
power level%Display the remaining battery level of the sensor as a percentage.
Before arranging sensors on site, it is necessary to perform equipment pairing operations to confirm that all sensors to be used are working properly. The specific steps are as follows:
1Open the tablet to the main interface, and upon entering the interface, the instrument will select an unused wireless channel frequency;
2Enter the device pairing interface and check all the options to be usedbiography
Sensor, ensuring that 'continuous reading of battery level' is in the selected state; Turn on the power switches on all the sensors to be used one by one, and view the status indicator lights and battery percentage in the "Device Pairing" interface.
3If the status and battery display of all sensors used are normal, return to the main interface and open the data zero calibration interface.
4According to the data zeroing steps, calibrate the necessary sensors (inlet pressure, outlet pressure, and temperature difference)),After the calibration is completed, the power of the tablet and sensors can be turned off.
5Install the on-site sensors, and after all the sensors to be used are installed, turn on the switches of the tablet and sensors again, and follow the testing steps for testing.
When there is abnormal sensor communication on the testing site, such asHow to handle:
1If the status of a sensor is displayed in gray, it indicates that there is a problem with the sensor. Please turn on the switch of the sensor again and try again.
2、If the status of all sensors is displayed in gray, it indicates that there is a malfunction on the tablet or that the tablet and sensors are not on the same channel. Please reopen the software and try again.
BCY-2A Wireless Multi parameter Tester for Mining Water Pump
1.1 Data zeroing interface
Click the 'Data Zero Calibration' button on the main interface to open the test zero calibrationThe interface is shown in the following figure:

Confirm before zero operation in school GPD1.5Wpressure sensor(Import pressure力)TheGPD20W Pressure sensor (outlet pressure)and GWD100W WenThe degree sensors have been successfully paired. Follow the prompts on the instrument to import and exportThe pressure probe is placed in the atmosphere, and the inlet temperature and temperature difference probes are tied together side by sideStable at the same water level 5 Minutes, the middle position of the interface will displayThe initial values of inlet pressure, outlet pressure, and temperature difference are displayed at the bottom of the interfaceThe zero value from the last zero calibration, when the displayed value is relatively stable, press the [OK] button,Perform zero calibration operation. Press the 【 Return 】 key to return to the main interface. Press the 'Next' button,Enter the input parameter interface for creating a new test.
1.1 Start testing interface
Click the 'Start Test' button on the main interface to open the 'Start Test' featureThe interface is shown in the following figure:

4Maintenance and Care
1. Measurement must be carried out under stable operation of the water pump.
2. The temperature sensing element is a platinum resistor, and a temperature measuring sleeve is used for temperature measurementThe material of the pipe is steel pipe or copper pipe, and oil should be injected inside the pipe to improve heat transfer or sealingSealing conditions.
3. The working temperature of the pressure probe is 70℃, when measuring pressure on site,Translate into EnglishThe pressure probe is installed on the joint with a three-way pressure pipe for measuring pressureWhen installing, you can first turn off the three-way valve or stop the water pump, install the probe, and then start it again).
4. Temperature and pressure probe components are easily damaged. During the storage period of the instrument after testing, it should be noted that the probe and cable connections should not be pulled or folded with force. They should be handled gently and not collided or dropped to avoid damage or affecting the accuracy of the instrument testingDegree.
5. The instrument cannot be disassembled by itself, and internal components should not be replaced or adjusted arbitrarily,To avoid human damage or measurement errors.
6. Please keep the instrument and probe clean during and after use; makeDuring use, transportation, and storage, attention should be paid to shock and moisture prevention, and it should be placed in a well ventilated and dry placeDry, indoor environment without corrosive gases around.
1 transportation and storage
1. The instrument power should be turned off during transportation and storage.
2. Do not store in high temperature places such as direct sunlight, carriages, or cabins.
3. Do not store in places with high humidity and dust.
2 Safety usage instructions
1. The use of instruments should comply with the relevant provisions of the Coal Mine Safety Regulations.
2. Users should carefully read the user manual and strictly follow itOperate the instrument.
3. During use, users are not allowed to open the instrument casing at will, let alone randomlyReplace the components of the instrument. The instrument has malfunctioned, please contact our unit.
4. Pay attention to waterproofing and prevent water from entering the instrument during storage and use.
5. Users must strictly follow the requirements of the mining measurement manual when using the instrumentConduct measurements.
6. During the storage period of the instrument, it is strictly prohibited to place it in high temperature, humidity, or corrosive conditionsThe location for storing sexual items.