●Automatic natural point measuring instrumentDesigned and manufactured according to the requirements of GBT 5332-2007 Test Method for Ignition Temperature of Flammable Liquids and Gases, used to determine the ignition temperature of flammable liquids and gases. This instrument adopts a temperature control method of *, with three-stage automatic temperature control and automatic measurement of the self ignition point of fire-resistant oil. It has the functions of automatic constant temperature, automatic sampling, automatic timing, automatic detection of self ignition point, automatic cleaning, and automatic printing of results after reaching the preset self ignition point.
Technical indicators
● Working power supply: AC220V ± 10% 50Hz
According to the standard: GBT 5332-2007 Test Method for Ignition Temperature of Combustible Liquids and Gases
● Display mode: 10 inch industrial grade color LCD touch screen, fully Chinese operation interface
Operating System: Embedded Windows system, software can be upgraded for free
Control system: Mitsubishi PLC system, with 16 built-in control programs
● Temperature control range: 80 ℃~800 ℃ Accuracy ± 1 ℃ Three stage temperature difference ≤ 1 ° C
Detection accuracy: meets standard requirements, repeatability ≤ 10 ℃, reproducibility ≤ 20 ℃
● Data input: Touch screen data input, which can input Chinese, English, numbers, punctuation, symbols, etc
Data output: Micro thermal printer, automatic printing, fast and clear printing
● Data Storage: Automatically stores 100 measurement results, allowing for viewing or printing of historical test data at any time
Instrument dimensions: length * width * height 550 × 500 × 700 (mm)
● Instrument weight: 60KG
● Environmental temperature: 5 ℃~30 ℃
Relative humidity: 30-80% RH
Technical indicators
●Automatic natural point measuring instrumentHigh degree of automation, automatic temperature control, automatic sampling, automatic timing, automatic detection of self ignition point, automatic cleaning, and automatic printing of results.
Adopting microcomputer control technology and embedded 10 inch color LCD touch screen, the operation is simple.
Equipped with a stepper motor, it has the characteristics of high precision, reliable operation, and long service life.
The digital detection method has high detection sensitivity and good repeatability.
Equipped with a built-in micro thermal printer that can automatically report detection results.
16 sets of measurement parameters can be preset for use in measuring different samples, facilitating measurement operations.
The preset parameters have the ability to be modified to meet the requirements for measuring special samples.
The design and manufacturing comply with national standards, and the instrument program can be upgraded when the national standards are revised.