- Phone
-
Address
8-9, Building B, Nanping Building, Nanping East Road, Chongqing
Chongqing Sainz Instrument Co., Ltd
8-9, Building B, Nanping Building, Nanping East Road, Chongqing
Monochromator type: Czerny Turner
Grating: 1800 lines/mm
Glitter wavelength of grating: 250nm
Wavelength range: 190nm~900nm
Spectral bandwidth: automatic switching between 0.1nm, 0.2nm, 0.4nm, 0.7nm, 1.0nm, and 2.0nm
Wavelength indication error: ± 0.30nm
Wavelength repeatability: ≤ 0.15nm
Absorption range: -0.1~2.5A
Dynamic baseline stability: ≤ 0.005A/30min (Cu spectral line)
Characteristic concentration: 0.02 μ g/mL/1% (Cu element)
Detection limit: 0.006 μ g/mL (Cu element)
Measurement repeatability:<0.7% (flame method, Cu element)
Lamp stand: 4 pieces
Host size: 700mm x 625mm x 570mm
Net weight: 80kg
Software functions:
Concentration calculation method: Standard curve method (single point calibration or multi-point calibration, multi-point calibration can choose 1st, 2nd, 3rd order functions and convert to each other) or standard addition method for testing, with a maximum of 20 repetitions, automatically calculating the mean, standard deviation (SD), relative standard deviation (RSD) and other functions.
Sample information table editing and data processing function: Provides sample information editing such as weight factor, volume factor, dilution factor, etc., which can directly calculate the actual sample concentration (sample content) and improve the completeness of experimental reports.