Welcome Customer !

Membership

Help

Ningbo Haier Xin Optoelectronics Technology Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Ningbo Haier Xin Optoelectronics Technology Co., Ltd

  • E-mail

  • Phone

  • Address

    Room 305, Building 1, Zhongchuang Technology Park, Jinyuan Road, Panhuo Street, Yinzhou District, Ningbo City, Zhejiang Province

Contact Now

Atmospheric methane/water vapor laser open circuit analyzer

NegotiableUpdate on 05/15
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The HT8600P atmospheric methane/water vapor laser open circuit analyzer adopts quantum cascade laser absorption spectroscopy technology, which uses high reflectivity mirrors exposed to the atmosphere on both sides to reflect mid infrared laser multiple times, with an effective optical path of tens of meters. It measures the weak absorption of mid infrared laser energy at the characteristic absorption peak of the target gas, and performs trace gas concentration inversion by real-time integration of the absorption peak spectrum curve. Its high-frequency concentration analysis characteristics make it very suitable for micro meteorological eddy related measurement technology. Combined with flux observation systems, it can accurately quantify the net exchange flux of methane/water between different ecosystems and the atmosphere.

Product Details

Core Technology

Quantum cascade laser technology:Using quantum cascade laser technology, utilizingMethane/WaterThe principle of molecules absorbing light energy under laser radiation of specific wavelengths, achieving the realization ofMethane/WaterNon contact measurement of concentration.

High resolution spectroscopy: This technology can obtain high-resolution spectral data, thereby improving the accuracy and sensitivity of measurements.

● Signal processing algorithm: The built-in signal processing algorithm can accurately analyze spectral data, convert it into methane/water concentration values, and eliminate the influence of interference factors.


Open optical cavity, ultra sensitive, fast response

● Portable design suitable for various on-site deployments

Suitable for areas without power grids and mobile platforms


Product model

HT8600P

Measure components

CH4

H2O

measurement range

0-15000 ppb

0-30000 ppm

Ensure accuracy range

1500-3000 ppb

0-30000 ppm

measurement accuracy

allan < 5 ppb

allan < 20 ppm

@0.1s @2000 ppb @STD

@0.1s @10000 ppm @STD

Default frequency for data output

10 Hz

ambient temperature

-10℃-45℃

ambient humidity

<99% R.H, without condensation

Atmospheric pressure range

70-110 kPa

power supply

12 VDC

system power consumption

~ 30 Watts (TYP)

Overall dimensions

Ø204 mm x1020 mm

weight

12 kg

contact method

RS232serial port

storage method

throughPCOr third-party data collectors

user interface

Windowssoftware

Optional functional modules

Automatic cleaning, rain sensing, lens heating


Environmental monitoring:Real time monitoring of methane/water in urban and industrial areas to assess air quality.

● Agricultural monitoring:Monitoring agricultural activitiesMethane/WaterOptimize fertilization strategies to reduce emissions and minimize environmental impact.

Industrial emission monitoring: used for monitoring and controlling industrial emissions of methane/water to ensure compliance with emission standards.