Aqua TROLL 100 and 200 CTD Data RecorderProduct Introduction
Aqua TROLL 200Measurable and recordable water level/Pressure, temperature, and conductivity,Aqua TROLL100 can measure and record temperature and conductivity. The conductivity cell design significantly reduces the instrument diameter (1.83cm) while ensuring measurement accuracy.Aqua TROLL100200Supports self recording mode, long-term online monitoring, and portable measurement, providing multiple solutions.
Applied to coastal, tidal, wetland, and estuarine research, flood and storm surge monitoring, site restoration and mine monitoring, groundwater and surface water monitoring.
Aqua TROLL 100 and 200 CTD Data RecorderProduct Features
·Real time salinity compensation technology is used to accurately measure data even in environments with changing salinity values.
·The sensor is calibrated through 3D and NIST traceable factory calibration within the complete pressure and temperature range. The gauge pressure (ventilation) system can meet the requirements of accurate height measurement.
·High data accuracy and small drift make it very suitable for long-term online monitoring.
·Supports multiple communication protocols such as Modbus/RS485, SDI-12, and 4-20mA. By connecting CTD to telemetry systems, SCADA, or HydroVu data service platforms, data can be accessed and notifications can be received at any time.
·The VuSitu Mobile app and Win-Situ5 PC software make it easy to configure instruments, generate reports automatically, simplify data management and calibration processes.
·Titanium metal structure is corrosion-resistant and pollution resistant, suitable for various harsh environments such as coastal areas, mine monitoring, and site restoration.
·A diameter of 1.83cm is suitable for various narrow monitoring environments.
·RuggedTable Systems quickly and securely connects with titanium twist lock connectors. The Rugged Cable Spliter can connect up to 8 In Situ instruments simultaneously, enabling multi layer deployment and same layer multi parameter deployment.
Technical Specifications
Aqua TROLL 100 and 200 | |||
temperaturescope |
Operating temperature:- 5 - 50 ℃ ( 23-122 ℉) Storage temperature:- 40-65 ℃( - 40 -149 ℉) |
||
size&Weight |
diameter: 1.83cm (0.72in.) Length: 31.5cm (12.4in.) Weight:188g(0.41lb) |
||
material |
Titaniummetalmain body,Delrinprotective cap, PVCConductivity pool |
||
communication protocol |
Modbus / RS485, SDI-12, 4 – 20 mA |
||
Battery Typeandlifespan |
3.6 V lithium batterypool;5 years or 200000areading |
||
外部power supplypower supplyAnd power consumption |
8-36 VDCMeasure current:15mA; Sleep current: 40uA |
||
memory |
4.0 MB |
||
data recording |
190,000 |
||
日志quantity |
50 |
||
Record mode |
Linear,Mean linearThe event |
||
fastestRecord speed |
linear: 1 time per minute, average linearity: 1 time per minute, event: 1 time per second |
||
fastestoutput rate |
1 second |
||
TriE Conductivity Sensor | |||
measurement method |
EPA Method 120.1; Standard Methods 2510 |
||
Range, accuracy,resolution |
scope:0- 100,000 µS / cm precision:080000 µ S/cm, ± 0.5% of reading+1 µ S/cm 80 ,000-100000µ S/cm, ± 1.0% of reading resolution: 0.1 µS / cm |
||
parameter |
range |
unit |
|
Actual conductivity |
0-100,000 µS/cm |
µS/cm, mS/cm |
|
Specific conductivity |
5-100,000 µS/cm |
µS/cm,mS/cm |
|
salinity |
0-42 PSU |
PSU |
|
溶解性固体TDS |
0-82ppt |
ppt,ppm |
|
resistivity |
10-200000 Ohms-cm |
Ohms-cm |
|
density |
0.98-1.14 g/cm3 |
g/cm3 |
|
pressure/Water level sensor | |||
range |
Absolute pressure(notventilation) 30 psia: 11 m (35ft) 100 psia: 60 m (197ft) 300 psia: 200 m (658ft)
|
gauge pressure(Ventilation) 5psig: 3.5 m (11.5ft) 15psig: 11 m (35ft) 30psig: 21 m (69ft) 100psig: 70 m (231ft) 300psig: 210 m (692ft) |
|
Swelling and crackingpressure |
Maximum pressure capacity twice the range; Expansion and cracking 3 times the range |
||
precision&Resolution |
Full range±0.05%Or better Resolution:Full range±0.01%Or better |
||
unit of measurement |
pressure: psi , kPa , bar , mbar , mmHg , inHg , cmH2 O , inH2 O . Water level:in,ft,mm cm m |
||
temperature sensor | |||
measurementmethod |
EPA Method 170.1 |
||
precision、resolution |
precision: ± 0.1 ℃ Resolution: 0.01 ℃ or better |
||
unit |
Celsius、Fahrenheit |
||