The adjustable airflow constant temperature aging chamber is used to test the performance indicators of components and materials of electronic and electrical products, automobiles and motorcycles, rubber, plastics, metals, marine weapons, universities, research institutions and other related products under high temperature and constant temperature changes.
1、 Box structure:
There are two glass doors that can be easily opened and closed. Large tempered glass observation window, providing clear visibility of the internal conditions at all times. Box insulation adopts**Fine glass fiber insulation cotton can avoid unnecessary energy loss.There is a distribution cabinet on the right side of the box, which can be used for external testing power lines or signal lines. The time and frequency of ventilation can be set arbitrarily. The test stand has multiple layers and can be freely adjusted in height.
2、 Technical parameters(Non standard sizes can be customized)
model studio dimensionsMm Overall dimensions mm
KM-HW-100B 800×1900×1600 900×2400×2200
3、 Performance indicators:
1. Temperature range: RT+10 ℃~60 ℃
2. Uniformity: ± 2 ℃
3. Constant temperature fluctuation: ± 0.5 ℃
4. Heating rate: about 3 ℃/min
5. Ventilation time: 1min to 9999h, adjustable temperature operation control system controller
6. Imported LED digital display P, I, D S, S, R. microcomputer integrated controller
7. Setting accuracy: Temperature 0.1 ℃
8. Indication accuracy: Temperature 0.1 ℃,
9. Resolution: 0.1 ℃
10. Heating system fully independent system, nickel chromium alloy electric heating heater
11. Timing range: 1 minute to 9999 hours, adjustable cycle system, high temperature resistant and low-noise motor.
12. Multi blade centrifugal wind turbine
4、 Security protection
1. Power supply * * load, short circuit protection;
2. Safety grounding protection;
3. Temperature protection;
4. Time setting protection;
5. Overload protection for blower motor;
6. To protect the equipment, all alarms will automatically cut off the power and sound a warning
5、 Precautions for use:
1. Power supply equipment: A * * load fuse device should be installed in the power supply line for this special box, and it should have a good grounding device.
2. Preparation of test samples: After numbering the test samples one by one, place them on the test sample rack, avoiding contact or collision with each other.
3. When everything is ready, close the box door.
4. The instrument casing must be effectively grounded to ensure safe use.
5. The instrument should be placed in a well ventilated indoor horizontal position, and no flammable or explosive materials should be placed around it.
6. An iron shell knife switch should be installed in the power supply line for this special box. Before grounding and energizing the shell, please check the electrical performance of this box and pay attention to whether there is any open circuit or leakage. This instrument has no explosion-proof device and should not be placed in flammable and explosive materials for drying.
7. The items inside the box should not be placed too tightly, and space must be left to facilitate the circulation of hot air.