ASTM D5453 Determination of Total Sulfur Content by Ultraviolet Fluorescence Spectroscopy

NegotiableUpdate on 07/07
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The UV fluorescence sulfur analyzer adopts the principle of UV fluorescence method and is suitable for determining the total sulfur content in crude oil, distillate oil, liquefied petroleum gas, plastics, petrochemical products, food, etc. After being introduced into the high-temperature cracking furnace, the sample undergoes cracking oxidation reaction. At a high temperature of around 1000 ℃, the sample is vaporized and undergoes oxidative cracking. The total sulfur content is determined by ASTM D5453 ultraviolet fluorescence spectrophotometry

Product Details

ASTM D5453 Determination of Total Sulfur Content by Ultraviolet Fluorescence Spectroscopy

ASTM D5453 Determination of Total Sulfur Content by Ultraviolet Fluorescence Spectroscopy

UV fluorescencemeasureSulfur analyzerPrinciple of determination using ultraviolet fluorescence methodSuitable for determining the total sulfur content in crude oil, distillate oil, liquefied petroleum gas, plastics, petrochemical products, food, etc. After being introduced into the high-temperature cracking furnace, the sample undergoes cracking oxidation reaction. in1000At high temperatures, the sample is vaporized and undergoes oxidative cracking, with sulfides quantitatively converted to sulfur dioxide. The reaction gas is carried by the carrier gas, and after passing through a membrane dryer to remove its moisture, it enters the reaction chamber. Sulfur dioxide is exposed to specific wavelengths of ultraviolet radiation, which absorbs this radiation and causes some electrons to shift to high-energy orbitals. Once electrons return to their original orbits, excess energy is released in the form of light and detected by a photomultiplier tube at a specific wavelength. The emitted fluorescence is specific to sulfur and proportional to the sulfur content in the original sample. After amplification by a microcurrent amplifier and computer data processing, it can be converted into an electrical signal proportional to the intensity of light. By measuring its size, the sulfur content of the corresponding sample can be calculated.

SH0689UV fluorescence sulfur analyzerconform toASTM D5453、SH/T 0689-2000Standard

GB/T11060.8 Natural gas - Determination of sulfur-containing compounds - Part 8: Determination of total sulfur content by UV fluorescence spectrophotometry

Main Features

Computer control operation can achieve human-machine dialogue, and data and peak shapes can be automatically stored and printed.

The operating software can be integrated with LIMISConnect the system to achieve automatic data transmission.

The instrument adopts a new type of quartz cracking tube, which can increase the injection volume and reduce measurement errors.

The automatic on/off of the cooling fan eliminates the hassle of waiting for the sample to shut down, saving time and effort.

The darkroom system adopts new foreign technologies, and key components use imported detection elements, which improves the detection accuracy of the instrument

The instrument adopts an imported membrane dryer to absorb moisture, saving the trouble of frequently replacing magnesium perchlorate and reducing the problem of air leakage.

The use of computer-controlled temperature has improved the temperature control accuracy of the instrument and is less prone to overheating.

Using air cooling instead of water cooling saves water resources.

Existing calibration curves can be retrieved and calibrated with standard samples for sample analysis.

Technical parameters

Measurement range: 0.1 ~10000mg/Lppm

Detection limit: 0.1mg/Lppm(Sulfur element)

Repetitive error:1.0mg/L<X≤10mg/L, Cv≤±10%

10mg/L<X Cv≤±5%

Gas source requirements: High purity oxygen:99.98%above;High purity argon gas: 99.98%Abovebring your own

Temperature control range: Room temperature~1200

Temperature control accuracy: ±1

Power supply AC220V±10% 50HZ

Environmental Requirements: temperature15~40; Humidity≤85%

Overall dimensions: Host:550×560×200(mm) Weight:9kg

temperature Control: 550×560×400(mm) Weight:10kg

● Packaging size: 705 * 560 * 545mm Weight: 40.32kg Wooden crate

● Packaging size: 705 * 600 * 555mm Weight: 42.10kg Wooden crate

Note: The appearance pictures, dimensions, volume, and weight of the instrument are for reference only. If there are any updates to the instrument, there may be slight changes without prior notice.