Thermo Fisher TLD Model 5500 Readers Thermoluminescence Readers

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

The Thermo Fisher TLD Model 5500 Readers thermal luminescence reader can cost effectively measure the radiation dose absorbed by TLD personal dosimeters. A single load can automatically process up to 50 TLD dosage components. Harshaw TLD Model 5500 Readers Thermoluminescence Reader | Thermoluminescence Dose Reader | Thermoluminescence Dose Meter | TLD Dose Meter | TLD Radiation Meter | TLD Dose Meter

Product Details

Thermo Scientific ™ Harshaw TLD Model 5500 Readers Thermoluminescence Readers can cost effectively measure the radiation dose absorbed by TLD personal dosimeters. A single load can automatically process up to 50 TLD dosage components. Harshaw TLD Model 5500 Readers Thermoluminescence Reader | Thermoluminescence Dose Reader | Thermoluminescence Dose Meter | TLD Dose Meter | TLD Radiation Meter | TLD Dose Meter|Thermo Fisher TLD Model 5500 Readers Thermoluminescence Readers

description

The compact and popular reading device features a linear, adjustable heating system and cooling photomultiplier tube, which can be combined with electronic components to measure the thermal luminescence output. Running WinREMS software on a Du Li computer provides a user interface, reader control, and application software.

  • Easy to operate, service, and maintain, with measurement quality assurance of up to50 of themThe unmanned automatic operation thermoelectric PMT cooler of the dosimeter provides multiple, adjustable, linear time temperature curves with high gain stability and automatic background subtraction functions; The heating curve includes preheating, acquisition, and annealing cycles using a linear non-contact hot gas heating method, with a temperature range of up to 600 ° C (1112 ° F) and dynamic acquisition of 7 orders of magnitude every 30 seconds. Normal TTP can withstand falling from a height of 20 millimeters onto cement ground. The ambient lighting can withstand up to 1000 lux of light (covered by the reader), including computer-processed deconvolution luminescence curve software, neutral filter, and calibration software

    Dose measurement performance when using LiF: Mg, Ti (TLD-100):

  • Organizational equivalence: approximate organizational equivalence
  • Batch uniformity: ± Batch uniformity standard deviation of 15%, can be reduced when special requirements are met. Residual TL signal:<0.2% when not annealed. Sensitivity change<10% for each dosimeter after more than 500 measurements. Threshold:<10 μ Gy (1 mrad). One unirradiated dosimeter is repeated 10 times, based on 2.26 standard deviations

Optical detection system

  • Dynamic range: Dynamic range: 7 orders of magnitude Preheating time: 30 minutes<linear deviation less than 1%:<stability: continuous reading 10 times, one standard deviation better than 1.0 mGy:<dark current: 50 mGy (relative to Cs-137) Test light: temperature stability, LED stability<deviation of 0.5% from continuous 10 readings, color: blue (wavelength 470 nm)

Dosimeter heating system

  • Method: Linear Thermal Nitrogen Time/Temperature Curve (TTP) Temperature Reproducibility: (± 1 ° C)

Computer Requirements (* Low): Intel ™ Pentium ™- Compatible with computers, 120MHz; 64MB RAM; 4GB hard drive; VGA color display, 800 × 600 * low resolution; Mouse and keyboard; CD-ROM drive; One RS232 serial port for connecting to the 5500 model; A parallel port for printers; Windows ™ operating system

Including: automatic sample changer and carrier disk

Thermo Fisher TLD Model 5500 Readers Thermoluminescence Readers

*Applicable to:

  • Radiation therapy | Validation of treatment plans | Radiation resistance validation Whole body irradiation dose validation Skin irradiation dose validation Stereoscopic radiation output factor measurement Dose validation of important organs Diagnostic dose research Quality assurance CT dose measurement Environmental dose determination Radiation food testing Radiochronological determination Breast | Room X-ray photography