Laboratory high-pressure reactorThe stirring bearing adopts corrosion-resistant and turbid stainless steel bearings, suitable for high-speed rotation and mixing of low viscosity materials. There are two discharge methods: upper discharge method and lower discharge method. The heating methods include electric heating, hot oil circulation, jacket steam heating, etc.
Laboratory high-pressure reactorIt is a typical innovation of magnetic transmission device applied to large-scale reaction equipment. It fundamentally solves the problem of shaft seal leakage that cannot be overcome by previous packing seals and mechanical seals, without any leakage or pollution. It is currently the most ideal device for chemical reactions under high temperature and high pressure in China, especially for chemical reactions of flammable, explosive, and toxic media, which further demonstrates its effectiveness. For the inner surface of the kettle with polishing requirements, the above mirror polishing level can be achieved. For high viscosity materials, it can be processed into a conical bottom for easy feeding and cleaning.
Laboratory high-pressure reactorPrecautions for Attention
1). The experimental autoclave should be used at the designated location and operated according to the instructions.
2). Identify the test pressure, usage pressure, and maximum usage temperature conditions engraved on the main container, and use them within their allowable range.
3). The pressure used by the pressure gauge should be within 1/2 of its indicated pressure. And often compare the pressure gauge with the standard pressure gauge and calibrate it.
4). Oxygen pressure gauges should be avoided from being mixed with pressure gauges used for other gases.
5). Safety valves and other safety devices should use instruments that have undergone regular inspections and meet the specified requirements.
6). Attention must be paid during operation to accurately insert the thermometer into the reaction solution.
7). The interior and lining parts of the high-pressure vessel should be kept clean.
8). When covering the disc flange cover, the bolts located on the diagonal should be tightened in pairs in sequence.
9). When the measuring instrument ruptures, in most cases, it breaks on both sides of its glass surface. Therefore, do not stand in these dangerous places during operation. When danger is expected, the glass should be removed and replaced with a new one.