Automatic natural gas analyzer operation video

Automatic natural gas analyzer fully automatic operation video

LNG tanker sampling video

Operation steps of natural gas analyzer:

  1. Turn on the power switch of the hydrogen generator, wait for the pressure of the instrument's' carrier gas I 'to reach the set pressure (usually 0.1-0.15Mpa), turn on the instrument's power, and turn on the' bridge flow 'switch;
  2. Open the computer (long press the power button) and open the calorific value version software;
  3. Connect the sampling bag for gas to the 'sample inlet' of the instrument, press the 'start' button, and the instrument will run according to the set program (automatic sampling, automatic injection, automatic signal collection). After 8 minutes, check the test results (nitrogen, methane, ethane, propane content, calorific value, density, white number, combustion potential automatically calculated).