Automatic natural gas analysis gas chromatograph operation video
Gas chromatograph operation
Power on/off of natural gas analyzer
1. Turn off the "bridge flow" switch (flat is off)
2. Turn off the instrument power, turn off the workstation and computer, and then turn off the hydrogen generator switch
Precautions for maintenance of natural gas analyzer:
1. Check the water level of the hydrogen generator and replenish it promptly if it is missing (between the upper and lower positions)
2. The color changing silicone of the hydrogen generator gradually changes from blue to pink. When two-thirds of the volume of the drying tube turns pink, a new silicone needs to be replaced or the original silicone needs to be baked in an oven (120-150 degrees) for 30 minutes to change color from pink to blue.