D&S AE1/RD1 emissivity measuring instrument: measures the surface emissivity of small or irregular materials
Date: 2025-09-22Read: 0

D&S AE1/RD1 emissivity measuring instrument:Test the emissivity of small or irregular material surfaces


The combination of D&S's AE1 emissivity meter and RD1 adjustable digital voltmeter provides a specially designed, efficient, and reliable solution to address the measurement challenges mentioned aboveIt combines laboratory level precision with the convenience of on-site applications.

image.png


Core Technology and Working Principle


AE1 emissivity meter is used for measurementTotal hemispherical emissivityThis is exactly the physical quantity required for most heat transfer calculations and standard complianceThe core innovation of its design lies in its unique working principle:


The detector inside the instrument is electrically heated to a stable temperature, while the sample itself does not need to be heated .


This design cleverly utilizes the principle of calorimetry measurement - determining emissivity by directly measuring the radiative heat exchange between the detector and the sample - but greatly simplifies the operation processTraditional methods typically require heating the sample, which is not only time-consuming but may also cause irreversible damage to the sample. The heating detector design of AE1 ensures non-destructive measurement and eliminates the need for any complex temperature measurements, revolutionizing the convenience and speed of measurement.

image.png


Key features and user benefits


  • Compliance and high precisionThis instrument complies withASTM C1371andISORelevant standards ensure the traceability of measurement results, which can be directly used for regulatory compliance and product certification. Its


    $pm 0.01$The repeatability and linearity of emissivity units provide users with highly reliable data support.


  • Simplicity and time-savingThe operation process of AE1/RD1 is designed to be extremely simple, requiring only two steps to complete the measurement:

    1. Place the AE1 probe on the accompanying high emissivity standard sample, rotate the gain knob on the RD1 voltmeter to make the reading consistent with the known emissivity value of the standard sample, and complete the calibration.

    2. Move the detection head to the surface of the sample to be tested, and the emissivity value displayed on the RD1 voltmeter is the emissivity value of the sample, which can be directly read.


      After about 30 minutes of preheating, each measurement only takes 90 seconds, which makes it very suitable for quality control sampling on production lines or large-scale inspections on construction sites.


  • Responding to diverse application demands:

    • Under standard configuration, AE1 can measure flat surfaces with a diameter of no less than 2.25 inches (5.7 cm). To meet diverse application needs, D&S offers a range of optional adapters:


    • AE-AD1andAE-AD3Adapter: It can reduce the minimum measurement area to 1.5 inches (3.8 cm) and 1.0 inch (2.54 cm) respectively, making it very suitable for testing small components or material samples.

    • AE-ADPAdapter: This is a key accessory designed specifically for measuring "difficult" surfaces. It can reliably measure low thermal conductivity materials (such as certain building insulation materials), rough or textured surfaces (such as roof tiles, coatings), and even large curvature cylindrical surfaces with a radius greater than 2 inches (which can be applied to coating inspection of automotive exhaust pipes or industrial pipelines).

    • Consumer adaptersD&S can also customize adapters for almost any geometric shape surface design according to customers' special needs, fully demonstrating its technical flexibility and problem-solving ability.


  • Portability and on-site applicationThe optional lithium-ion battery pack can provide up to 12 hours of continuous working time and is equipped with a low battery indicator function. This option transforms AE1/RD1 from a desktop instrument into a truly portable on-site tool that can easily handle both on-site inspections of building roofs and inspections of spacecraft components in aircraft hangars.

    image.png


Table: Key Technical Specifications of D&S Model AE1/RD1


The following table summarizes the core technical parameters of AE1/RD1 emissivity meters, providing a quick reference for technicians.

Specifications Value/Description
Measuring principle Approximate measurement of total hemispherical emissivity at 65 ° C (150 ° F)
meet the standard ASTM C1371, ISO requirements
repetitiveness $pm 0.01$Emissivity unit
linearity $pm 0.01$Emissivity unit
usability Detector electric heating (sample not heated); No temperature measurement required
Measuring speed After 30 minutes of preheating, read approximately 90 seconds per time
Standard sample size 2.25 inches (5.7 cm) flat surface
Optional adapter Suitable for flat samples as small as 1.0 inch (2.54 cm); Suitable for rough, low thermal conductivity, or cylindrical surfaces
reading device RD1 type adjustable digital voltmeter, directly reading emissivity
Attached Attachment High/low emissivity standard samples, heat dissipation blocks, carrying case, universal power supply
Key options Battery pack (12 hour battery life), ISO certificate, 10x gain option (for readings with 3 decimal places accuracy)