Dissolved oxygen analyzerThe operation steps are as follows:
Preparation before use
Electrode polarization: If the instrument has not been used for a long time, the electrode should be connected to the host and polarized for 20-30 minutes to ensure the stability of the sensor.
Probe inspection:
Check if the probe film is intact, without damage or aging;
Confirm that the metal electrode is tightly attached to the film and free from rust, and if necessary, use fine sandpaper to polish it;
Check if the electrolyte inside the probe is sufficient. If there are bubbles, replace the electrolyte with a new one (it is recommended to use the original electrolyte).
Reagent preparation: Prepare reagents according to calibration requirements, such as 5% freshly prepared sodium sulfite solution (zero oxygen calibration solution), deionized water, etc.
Water sample collection:
Collect water samples using a clean and odorless sampler to avoid contamination;
It is recommended to collect at a location close to 50cm underwater to ensure the representativeness of the water sample;
If there is a significant difference between the temperature of the water sample and room temperature, it should be allowed to stand until the temperature stabilizes before measurement.
instrument calibration
Zero oxygen calibration:
Immerse the electrode in zero oxygen calibration solution (5% sodium sulfite solution) and gently stir;
After the reading is stable, adjust the zero adjustment knob or relevant settings to make the instrument display 0.
Full oxygen calibration:
Clean the electrode and dry it with filter paper;
Place the electrode close to the surface of a conical flask containing deionized water (without contact with water) and wait for the reading to stabilize;
Adjust the span calibration knob or related settings to indicate the saturated dissolved oxygen value of pure water at this temperature.
Attention: The calibration steps for different models of instruments may vary slightly, and specific operations should refer to the instructions.
Dissolved oxygen analyzerMeasurement operation
Electrode insertion: Immerse the electrode into the water sample to be tested, ensuring sufficient contact with the water sample.
Mixing water sample:
Gently stir the water sample with a mixer or manually, keeping the probe always in the water sample;
To avoid measurement errors caused by oxygen consumption near the probe.
Wait for stability: According to the measurement principle of the instrument, wait for 5-10 minutes (or the instrument prompts a stable reading) before reading the data.
Read data:
Record the dissolved oxygen concentration value displayed by the instrument (unit: mg/L);
Simultaneously record information such as measurement date, time, and water sample temperature for subsequent analysis.
Post use maintenance
Electrode cleaning:
After use, rinse the electrode with deionized water to remove any residual water sample;
Avoid wiping the electrodes with hard objects to prevent damage to the film.
Electrode storage:
Short term idle: immerse the electrode in deionized water to keep the probe moist;
Long term idle: Separate the probe from the host and store the electrodes in a dry environment.
Instrument shutdown: Turn off the power of the instrument and remove the battery (if not used for a long time) to avoid battery leakage and damage to the instrument.
Regular calibration: It is recommended to calibrate the instrument every 10 days to ensure measurement accuracy; If the instrument is used frequently, the calibration cycle can be shortened.
Consumables replacement: Regularly replace consumables such as electrolytes and probe films, with specific cycles determined by usage frequency and manufacturer recommendations.