Technical parameters of dissolved oxygen analyzer
Date: 2017-04-13Read: 28
PH:-2.00.....+16.00pH
mV:-1999.....+1999mV
Accuracy:
PH:±0.01
MV: ± 0.3mV (+15...+35 ℃) Temperature: -5.0 ℃...+105.0 ℃
Temperature compensation: TFK325/Pt1000 temperature probe or manual compensation
Display: LCD display, 60 × 45mm, simultaneously displaying temperature, pH value, and other characters such as electrode performance and electrode status
Electrode testing: Determine electrode performance based on calibration results and display on LCD
Correction alarm: adjustable from 1 to 999 days
Automatic reading: making reproducibility better than 0.02PH
Correction: MultiCal? Automatic calibration, AutoCal: Automatic 2 or 3 point calibration using DIN standard solution, AutoCal TEC: Automatic 2 or 3 point calibration using WTW standard solution, ConCal: Normal 2 point calibration, standard solution optional
Calibration Protocol: GLP Calibration Protocol, Instrument Storage Latest Calibration Protocol
O2 concentration: 0.00.... 19.99mg/l
0.0.....90.0mg/l
Saturation: 0.0... 19.99%
0.....600%
Temperature: -5.0....+50.0 ℃
Accuracy:
Dissolved oxygen analyzer
Dissolved oxygen: ± 0.5% test value
Saturation: ± 0.5% test value
Temperature: ± 0.1K
Pressure compensation: automatic compensation, built-in pressure probe (500.... 1100hPa)
Salinity correction: automatic, range 0.0... 70.0, * adjustable
Temperature compensation: automatic
Display: LCD display, 60 × 45mm, simultaneously displaying temperature, dissolved oxygen, and other characters such as electrode performance and electrode status
Electrode testing: Determine electrode performance based on calibration results and display on LCD
Correction alarm: adjustable from 1-99 days
Automatic reading: making reproducibility better than 0.05mg/l
Calibration: Conducted in OxiCal (r) calibration kit, with constant pressure and temperature, accurate and reliable calibration
Data storage: It can store 800 sets of data (dissolved oxygen, temperature, date/time, sample number), with adjustable storage intervals between 5-60 seconds, and data storage is not easily lost
Data storage: automatic storage
Output: RS232 interface