controlArabidopsis thaliana incubatorIt is a multi-purpose product designed for studying Arabidopsis plant experiments, which can adjust the layering as needed, quickly change the experimental conditions required for research, and achieve a high level of flexibility, accuracy, and versatility.
The humidity control of Arabidopsis thaliana incubators needs to be combined with ventilation systems and humidity regulation devices to accurately regulate according to different growth stages. Here are the specific methods:
Principle of humidity control:
Sensor monitoring:
Real time monitoring of indoor humidity through humidity sensors, with data feedback to the control system.
Adjustment of execution mechanism:
Based on sensor data, automatically activate the humidifier (increase humidity) or dehumidifier (decrease humidity).
Staged humidity control:
Seedling stage (95% humidity):
Close the exhaust outlet, reduce air exchange, and maintain a high humidity environment.
If the humidity is too low, it can be replenished with moisture through a humidifier.
Maturity period (80% humidity):
Turn on the exhaust outlet or fan to promote air exchange between the inside and outside of the box and reduce humidity.
Use a dehumidifier to assist in control and avoid excessive humidity affecting plant transpiration.
Ventilation system settings:
Intermittent operation: Start the fan in a short cycle (such as running for 5 minutes every 30 minutes) to avoid excessive humidity fluctuations caused by continuous ventilation.
Wind speed regulation: Reduce the fan speed to minimize the interference of airflow on humidity.