Fruit fly incubatorA constant temperature and humidity experimental equipment mainly used for insect genome analysis, gene mutation sequencing, and other purposes using fruit flies as model organisms. Suitable for research in life sciences, cellular microbiology, molecular genetics, etc; It is an ideal auxiliary equipment commonly used in scientific research and production in many fields such as biotechnology, medicine, agriculture, aquaculture, and animal husbandry, providing precise constant temperature and humidity experimental conditions.
Good design of circulating air ducts, air duct ventilation structure, and equal return air technology; The studio has gentle wind speed and uniform temperature and humidity (2017).
Microcomputer automatic control; LCD display, touch switch, easy to operate.
Equipped with power-off memory function, it can continue to operate according to the original set value from the power-off point when powered on again.
Usage guidelines:
Environmental control:
Temperature: 22-25 ℃ (± 1 ℃ fluctuation)
Humidity: 40-60%, to prevent mold growth
Light: 12 hour day night alternation cycle
Operation process:
Check the temperature and humidity sensor before placing the fruit fly
Avoid frequent opening of doors, with each operation lasting no more than 30 seconds
Record abnormal environmental parameters after use
Cleaning and disinfection:
Daily maintenance:
Wipe the inner wall with 75% alcohol every week
Dismantle shelves, water trays, and other components for cleaning every month
Deep disinfection:
Using hydrogen peroxide dry mist sterilization technology during pollution
After disinfection, ventilate for 24 hours before use
Fault handling:
Abnormal temperature: Check the heating element/cooling system, calibrate the sensor
Humidity out of control: replace humidifier water tank, check sealing strip
Alarm handling: Immediately cut off power and contact professional maintenance
Safety precautions:
Personal protection:
Wear gloves and masks during operation
No eating or drinking in the experimental area
Equipment safety:
Stay away from vibration sources (such as centrifuges)
Regularly check the circuit for leakage prevention
Emergency response:
Transfer samples during contamination and thoroughly disinfect them
Flammable reagents need to be stored separately