Common faults and solutions of HALOSID pure water machine
Date: 2025-10-23Read: 2
The common faults and solutions of HALOSID pure water machine are as follows:
Fault phenomenon Possible reasons solution
High pressure pump does not start, unable to produce water Power outage or unplugged plug Check the power supply, ensure it is powered on and plugged in properly
Low voltage switch malfunction Check the low-voltage switch and replace it if it is damaged
High voltage switch or water level controller malfunction Check and repair or replace high-voltage switches and water level controllers
Transformer and fuse burned out Check transformers and fuses, replace damaged components
The high-pressure pump is working normally but cannot produce water High pressure pump loses pressure Check the high-pressure pump, repair or replace it
Pre filter clogged Replace the front filter element
Check valve blockage Check and clean or replace the check valve
Electromagnetic valve malfunction Check the solenoid valve and replace it if there is a malfunction
RO membrane blockage Clean or replace the RO membrane
High pressure pump does not stop Insufficient pressure of high-pressure pump Check the high-pressure pump, repair or replace it
Check valve blockage Check and clean or replace the check valve
High voltage switch malfunction Check and replace the high-voltage switch
Electromagnetic valve malfunction Check the solenoid valve and replace it if there is a malfunction
The high-pressure pump stops, but the wastewater does not stop Electromagnetic valve malfunction Check the solenoid valve and replace it if there is a malfunction
Check valve pressure relief Check and replace the check valve
After the water is full, the machine repeatedly jumps up Insufficient pressure of raw water Check the pressure of the raw water, and if it is insufficient, increase the pressure
Check valve pressure relief Check and replace the check valve
High voltage switch malfunction Check and replace the high-voltage switch
There is a pressure relief phenomenon in the system Check the system and repair the pressure relief point
The pressure tank is full of water, but pure water cannot flow out Pressure bucket pressure relief Inflate or replace the pressure drum
Rear filter clogged Replace the rear filter element
Pressure bucket ball valve damaged Check and replace the pressure drum ball valve
Insufficient pure water flow rate Pre filter clogged Replace the front filter element
Insufficient pressure of high-pressure pump Check the high-pressure pump, repair or replace it
RO membrane blockage Clean or replace the RO membrane
The wastewater ratio is imbalanced or the flushing switch is not tightly closed Check the wastewater ratio and flushing switch, adjust or turn off
The ratio of pure water to wastewater is severely imbalanced The flushing solenoid valve is damaged Check and replace the flushing solenoid valve
The wastewater proportioner is blocked or too conductive Check and clean or replace the wastewater proportioner
RO membrane blockage or severe decrease in removal rate Clean or replace the RO membrane
The machine keeps flushing Wastewater ratio failure Check and replace the wastewater ratio
The flushing switch is not tightly closed Turn off the flushing switch
loud noise The raw water pipeline is cut off, and the booster pump is idling Check the raw water pipeline to ensure water supply
There is gas in the water Check the condition of the raw water and eliminate any gas
Booster pump malfunction Check the booster pump, repair or replace it
The placement of the pure water machine is unstable Stable placement of pure water machine
The water pipe is too long, and the vibration of the pump causes the water pipe to strike the casing Adjust the position of the water pipe
Compressor failure Check the compressor, repair or replace it
water leakage Water leakage in the water circuit of the pure water machine Check the waterway, locate the leakage point and repair it
Small Unicom leakage Check and replace the small Unicom
Water dispenser leaking Check the water dispenser and repair the leakage point
Not producing pure water Pipeline problem of pure water machine Check and straighten out the pipeline
Power issue Check the power supply to ensure normal power supply
High voltage switch or float switch malfunction Check and repair or replace high-voltage switches and float switches
Inlet solenoid valve malfunction Check the inlet solenoid valve and replace it if there is a malfunction
Insufficient pressure after booster pump Check the booster pump, repair or replace it
RO membrane damage Replace RO membrane
Rear activated carbon filter clogged Replace the rear activated carbon filter element
TDS value too high The TDS value of the raw water is too high Understand the TDS value of raw water and determine whether it meets the requirements
The desalination rate of RO membrane decreases Check the RO membrane and replace it if it is damaged
Pure water stored in the water storage tank for too long Replace the water in the water storage tank
Rear activated carbon filter element not washed Rinse the post activated carbon filter element