Accurate determination of free antigens in 100% serum using Gyrolab
Date: 2025-11-05Read: 1


Accurately measuring the concentration of free antibodies/antigens in serum is crucial in drug development, especially in pharmacokinetic/pharmacodynamic (PK/PD) studies of monoclonal/multispecific antibody therapeutic drugs. Traditional methods often result in biased results due to sample dilution, prolonged incubation time, or unstable antibody antigen complexes.


Gregor Jordan et al. from Roche developed an immunoassay method for detecting free antigens in whole serum based on the Gyrolab platform. The method aims to address the challenges in free substance analysis, including:



The dilution of the sample leads to the dissociation of the antigen antibody complex,Overestimating the concentration of free antigen


The half-life of antibody antigen complexes is short (<100 seconds), and traditional LBA methods cannot accurately capture them


It is difficult to distinguish between free and bound antigens in the presence of therapeutic antibodies

使用Gyrolab实现100%血清中游离抗原的准确测定

This method adopts the principle of bridging immunoassay, and the specific steps are as follows:

1

Coat the labeled capture antibody onto a streptavidin affinity column

2

joinUndiluted serum sampleTo bind free antigens with captured antibodies

3

joinLabeled detection antibodiesBinding to another epitope of the antigen

4

Detection signal, quantitative free antigen

使用Gyrolab实现100%血清中游离抗原的准确测定

Key requirements of the method






Capture antibodies and detect antigen epitopes recognized by antibodiesnon-overlapping


Capture antibodies and therapeutic antibodiesCompete to bind to the same epitope of the antigen


Very short incubation time (preferably<2 seconds)


No need for sample dilution to avoid complex dissociation

使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定

Research Objective

Evaluate the "clearance efficiency" of antibody drugs on CCL2, that is, whether the antibody effectively binds and clears free CCL2.

使用Gyrolab实现100%血清中游离抗原的准确测定

研究method

Platform and Methods

使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定

Experimental Design

01

Calibration curve and quality control

使用Gyrolab实现100%血清中游离抗原的准确测定

Calibration range: 2430 pg/mL → 10 pg/mL

QC sample:

High QC:1820 pg/mL

Mid QC:230 pg/mL

Low QC:30 pg/mL

Acceptance criteria

± 20% deviation, CV < 15%

02

Drug intervention group

使用Gyrolab实现100%血清中游离抗原的准确测定

CNTO0888: Monospecific anti-CCL2 antibody (control)

CKLO2-SG1095: Bispecific anti-CCL2 antibody (experimental group)

Concentration setting:

Low dose: 7.5 ng/mL or 15 ng/mL

High dose: 10 μ g/mL

Results and Discussion

使用Gyrolab实现100%血清中游离抗原的准确测定


CNTO0888 almost completely eliminates free CCL2 at high concentrations


The clearance efficiency of CKLO2-SG1095 is slightly lower, but still significantly better thanblank group


At low doses, dual antibodies still have moderate clearance ability, indicating their dose-dependent nature




使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定

Research Objective

Quickly and accurately measure the concentration of free complement C5 in undiluted serum samples for evaluating the in vivo efficacy of anti-C5 antibody drugs such as eculizumab and crovalimab.

使用Gyrolab实现100%血清中游离抗原的准确测定

研究method

Platform and Methods

使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定

experimentresult


Linear range of standard curve: 0.55-3000ng/ml


LOQ:0.5 ng/mL, Capable of detecting<0.1% free C5(Total C5 ≈ 400 nM)

使用Gyrolab实现100%血清中游离抗原的准确测定
使用Gyrolab实现100%血清中游离抗原的准确测定

Summary

Advantages of Method and Technology

使用Gyrolab实现100%血清中游离抗原的准确测定

The Gyrolab platform has a no incubation flow-through mode,Greatly reduce non-specific binding in the matrix and effectively reduce matrix interferenceIn free substance analysis, it can supportWhole serum testingAnd alsoReduce the dissociation of antibody antigen complexes caused by incubationMake the analysis of free substances more accurate and efficient.






References:

[1] Method for determining the free antigen of an antibody in a sample, Gregor Jordan, Martin Schaefer, Maria Viert; Pub No. US2023/0393125 A1。