Summary of Troubleshooting Methods for Air Generator Malfunctions
Date: 2017-11-23Read: 20

Currently available in the marketair generatorGenerally, it has the function of delayed emptying after startup, which means that within the first 10 minutes of starting up, the air generator needs to exhaust the output gas to the atmosphere due to low gas purity and the presence of air in the pipeline system. Most manufacturers use a fixed gas barrier at the outlet of the exhaust valve to control the flow rate of the exhaust gas. This method can control the output gas flow rate during the exhaust process. However, when the exhaust is completed and the air switches to the chromatographic gas path, the output air quickly builds up pressure in the connected pipeline, causing the output flow rate of the air generator to increase rapidly. The electrolytic separation cell quickly separates the air, resulting in a large amount of unseparated air directly entering the chromatographic system, causing damage to the chromatographic column or rapid failure of the deoxygenation tube. Some manufacturers add a needle valve before the exhaust port to limit the flow. This method can cause additional problems. When the air system switches from exhaust to normal supply, the output flow rate of the needle valve will gradually decrease as the column head pressure of the chromatograph gradually rises and stabilizes. If you want to obtain normal flow, you need to adjust the needle valve diameter again, which will contaminate the stable high-purity air system again. To achieve high-purity and stable air, in addition to overcoming the above problems, it is also necessary to overcome the blockage and reflux phenomenon of the electrolytic separation cell.

  air generatorCommon faults and troubleshooting methods:
Simple faults and troubleshooting methods:
1. After turning on, the green working indicator light is on, and the compressor starts but the pressure does not increase. 1. If it starts * times after opening the box, please check the air inlet and replace the plug with a filter.
2. Block the output port, check if there is any air leakage at the joints of the drying tube and gas circuit, and seal the leaking area.
After turning on, the green working indicator light will turn on and the compressor will not start. Please open the right cover of the instrument and check if the thermal protection relay behind the compressor is burned out. If so, please replace it.
After turning on, the green working indicator light does not light up, the compressor does not start, and there is no output pressure. Please open the right cover of the instrument and check:
1. Is the transformer working properly;
2. 2. Are the pressure switches for 0.5 Mpa and 0.8 Mpa conductive? The conductivity is normal.
Replace faulty parts.
After starting up, it can work normally, but during the working process, the pressure drops below the set pressure, and the compressor does not start. Such faults are generally caused by thermal protection, and those that are prone to thermal protection include:
1. Air intake blockage (cleaning air filter)
2. Compressor oil shortage (add 200 grams of No.18 refrigeration oil).
3. One way valve leakage causes high starting pressure (replace the one-way valve).
The zero level air generator is designed specifically for gas chromatographs that require high-purity air, and is more suitable for use in environmental testing, scientific research, and other units. Without oilair generatorOn the basis of adding a hydrocarbon removal device, the use of high-temperature catalysts can reduce the concentration of hydrocarbons in compressed air to below 0.1ppm. The quality requirements of gas chromatography for auxiliary gas, the innovative independent high-temperature catalyst structure extends the service life of the machine, and the maintenance cost is lower compared to other brands.