Small knowledge on maintenance specifications for lithium battery low-voltage simulation testing machine
Date: 2025-07-17Read: 0
Lithium battery low voltage simulation testing machineAdopting a three-layer glass structure with steel mesh embedded in the middle to ensure safe visualization of the testing process; The outer box is made of cold-rolled steel plate and treated with excellent baking paint technology, which can effectively resist the influence of the external environment and ensure the long-term stable operation of the equipment.
By using a high-performance vacuum pump, it is possible to quickly reduce the pressure inside the box.
Support remote control start and stop, greatly improving the flexibility and efficiency of operation.
Support standard testing procedures and personalized testing solutions to meet various diverse testing needs.
The instrument is equipped with an automatic exhaust function and a manual pressure relief valve to ensure safe gas discharge inside the box.
Equipped with an excellent control system, real-time monitoring of air pressure and battery voltage changes inside the box, and automatic adjustment of the test cycle.
Color one key emergency stop button, which can directly cut off power in case of accidents, ensuring the safety of test personnel.
Built in explosion-proof lighting system ensures clear observation of the test status even at night or in low light conditions.
Equipped with an excellent control system, real-time monitoring of air pressure and battery voltage changes inside the box, and automatic adjustment of the test cycle.
Key points of operation and maintenance:
Testing process:
Sample drying treatment → placed in the center of the furnace → set pressure/time → start the vacuum pump → automatically record data;
It is strictly prohibited to open the box during testing, and samples must be taken after cooling to below 200 ℃.
Maintenance standards:
Clean the furnace oxide layer (soft bristle brush) every week to avoid damaging the heating elements;
Verify the vacuum gauge and thermocouple every six months, and replace them if the error is greater than ± 5 ℃;
The sealing strip is coated with high-temperature resistant silicone grease, and the closing pressure is maintained at 0.05-0.1MPa.