E-mail
mk@lubex.com.cn
-
Phone
18922417705
-
Address
2203, the Pearl River International Building, No. 112, Yuehua Road, Yuexiu District, Guangzhou
Guangzhou Lvbaicao Scientific Instrument Co., Ltd
mk@lubex.com.cn
18922417705
2203, the Pearl River International Building, No. 112, Yuehua Road, Yuexiu District, Guangzhou
Reference standards:2025 Pharmacopoeia - Determination Method for Natural Borneol (Dextrose) Content
Chromatography column: LBC-DEX3Cyclodextrin chiral columnColumn length: 30m, inner diameter: 0.25mm, film thickness: 0.25 μ m
Item number: ZDEX3-30-25-25,
Testing instrument: Agilent 8890, FID detector
柱温: 75℃ to 100℃ at 1℃/min, at 3℃/min to 150℃, at 10℃/min to 220℃(10min)
Vaporization temperature: 200 ℃
Detection temperature: 220 ℃
Column flow rate: 1.0ml/min
Split ratio: 5:1
Injection volume: 1uL
The main active ingredient in natural ice flakes is right-handed borneol, while the effects of left-handed substances are different and must be strictly distinguished.
The pharmacopoeia method uses a 30% seven - (2,3-O-dimethyl-6-O-tert butyldimethylsilyl) - β - cyclodextrin (14% cyanopropyl phenyl) methylsilicone capillary column (column length of 30m, inner diameter of 0.25mm, membrane thickness of 0.25 μ m) with high technical barriers.
The LUBEX DEX3 cyclodextrin chiral column adopts a derivatized cyclodextrin chiral stationary phase that matches the requirements of the pharmacopoeia. It has high recognition ability for chiral compounds and achieves baseline separation of levo - and right-handed Longines, which can meet the requirements of the pharmacopoeia.