Analysis of the Causes of Jaw Fracture in Steel Wire Pulling Machine
Date: 2025-11-21Read: 1


This is a very common and classic problem in material mechanics performance testing. The interruption of the steel wire in the tensile test at the clamp is mostly not due to insufficient strength of the steel wire itself, but rather caused byAbnormal fracture caused by stress concentration and clamping damage.

Simply put, it refers to the steel wire in the area near the jaws being clampedEarly destruction. The following are the detailed reasons and solutions that led to this problem:

Core reason: stress concentration

In an ideal tensile test, the stress in the gauge length section (the uniform middle part) of the specimen is uniformly distributed. But near the jaws, the situation becomes complicated:

1.Geometric shape mutation: The steel wire is suddenly clamped tightly by the jaws from a free state, and this clamping point is a mutation point of geometric shape and constraint conditions.

2.Stress concentration: All external forces are transmitted through this clamping point, resulting in local stress at the edge of the jaw being much higher than the average stress in other parts of the specimen. Even if the average stress has not yet reached the tensile strength of the steel wire, the peak stress at the edge of the clamp may have exceeded the material limit, leading to crack initiation and rapid fracture.

Specific cause analysis and solutions

1. The problem with the jaws themselves

Jaw type mismatch:

Problem: The wrong pliers were used. For example, using flat jaws to clamp round steel wires for testing flat specimens can result in line contact, high pressure, and easy damage and cutting of the steel wires.

Solution: Must use a shape that matches the sample shapeVType jaw.VThe type of clamp can better wrap round steel wire and connect the points/Line contact becomes surface contact, making stress distribution more uniform.

Jaw wear or damage:

Problem: After long-term use, the teeth of the jaws may be worn flat, with pits, chipped corners, or corroded. This will lead to a decrease in clamping force, requiring greater clamping force to prevent slipping, thereby exacerbating damage to the steel wire; Meanwhile, the damaged jaws themselves will generate many stress concentration points.

Solution: Regularly inspect and replace worn jaws. Ensure that the jaw teeth are clear and intact.

2. The problem of clamping operation

Inappropriate clamping force:

Problem: If the clamping force is too small, the steel wire will slip inside the jaws. Once it slips, it will cause serious scratches and compression on the surface, forming a crack source. Excessive clamping force can directly flatten or damage the steel wire, causing serious mechanical damage and stress concentration at the jaws.

Solution: Apply appropriate and uniform clamping force to just prevent slipping. For hydraulic or pneumatic clamping equipment, appropriate pressure can be set according to standards or experience.

Negative alignment:

Problem: The sample is not installed on the centerline of the fixture, resulting in eccentricity. This will result in the specimen being subjected to additional bending stress in addition to tensile stress. The bending stress is maximum at the edge of the jaw, greatly promoting the generation of cracks.

Solution: Carefully install the sample to ensure its alignment. For automatic centering fixtures, they should be allowed to adjust freely in place.

Insufficient clamping length:

Problem: The clamping length of the steel wire in the jaws is too short. This will result in insufficient constraint of the clamp on the steel wire, making it easy to slip, and the stress cannot be effectively transmitted smoothly from the clamp to the gauge length section, leading to a high concentration of stress at the outlet of the clamp.

Solution: Ensure sufficient clamping length, usually requiring clamping length not less than the diameter of the steel wire3-5Double, or follow the relevant testing standards.

3.The issue of sample preparation

Jaw damage:

Problem: During the clamping process, the teeth of the jaws have caused visible indentations, dents, or scratches on the surface of the steel wire. These damages themselves are tiny cracks that can directly become the starting point of fracture.

Solution:

Use jaws with finer and denser tooth patterns (for high-strength steel wires).

Adding padding to the steel wire clamping area is a very effective method. Common cushioning materials include:

Thin copper and aluminum sheets: They have good plasticity and can fill the gaps between the jaws and steel wires, dispersing pressure.

Sandpaper: Facing the sand surface towards the steel wire can increase friction and prevent slipping, while also serving as a buffer and dispersing pressure.

Special plastic gasket.

The sample itself has defects:

Problem: If the steel wire itself has original defects such as scratches, microcracks, or rust near the jaws, then this is naturally the weakest link.

Solution: Check the surface quality of the sample during sampling to avoid using sections with obvious defects as clamping parts.

Summary and troubleshooting steps

When encountering steel wires that always break in the jaws, it is recommended to investigate in the following order:

1.Firstly, check the jaws: whether they are correspondingVType of jaw? Are the teeth worn? Is it clean?

2.Check the clamping section of the sample: Is there any obvious indentation or damage on the surface? If there is, it indicates that the clamping force is too high or the jaws are not suitable.

3.Try using padding: This is a simple, low-cost, and often effective solution. Wrap the clamping part with thin copper sheet or sandpaper (with steel wire attached to the sand surface) and then clamp it tightly.

4.Check operation: Is the clamping force appropriate? Has the sample been aligned? Is the clamping length sufficient?

5.Observe the location of the fracture: If the fracture always occurs on one side of the jaw edge, it is likely a problem with poor alignment. If it is inside the jaws, it is more likely to be caused by excessive clamping force or jaw damage.

By systematically analyzing and adjusting the above factors, the vast majority can be basically solvedThe issue of "breakage in the jaws" is addressed in order to obtain test results that truly reflect the material's properties.