E-mail
shdalck@163.com
-
Phone
13611654820
-
Address
Building 3, No. 999 Dongye Road, Sijing Town, Songjiang District, Shanghai
Dalk (Shanghai) Instrument Equipment Co., Ltd
shdalck@163.com
13611654820
Building 3, No. 999 Dongye Road, Sijing Town, Songjiang District, Shanghai
1、 Core challenge: Why are ordinary electrodes unable to handle it?
There are three major destructive factors in high acidity and alkalinity working conditions:
Glass film dissolution: Strong alkaline solution (pH>12) can corrode sensitive glass films, resulting in delayed response and decreased slope
Acid error: High H ⁺ concentration in strong acid solution (pH<2) causes "acid error", and the reading deviates from the true value
Reference system poisoning: Ag/AgCl reference electrode forms insoluble substances when encountering sulfides, bromides, etc., blocking the liquid network
2、 Key selection elements: Four major technical indicators
1. Material of electrode membrane
Strong alkaline working condition: choose high lithium glass film (with 3-5 times increased alkali corrosion resistance)
Strong acid/HF containing conditions: Fluorine ion resistant glass film must be used (ordinary glass will be perforated in 1% HF for 24 hours)
Severe working condition: Antimony electrode (strong acid resistance, but slightly lower accuracy, ± 0.1pH)
2. Reference system
Open liquid network part: anti blocking design, suitable for suspended solids and viscous media
Double salt bridge structure: The outer salt bridge uses high concentration KCl+anti pollution electrolyte to delay reference poisoning
Solid polymer reference: Non porous design to prevent electrolyte contamination and blockage
3. Material of electrode body
Conventional corrosion: Polyether ether ketone (PEEK), resistant to organic solvents and weak acids and bases
Strong corrosion: Titanium alloy (Ti6Al4V) or Hastelloy, resistant to 98% sulfuric acid, 40%
Sanitary grade: 316L stainless steel, suitable for CIP/SIP cleaning in the food and pharmaceutical industries
4. Temperature and pressure
High temperature type: resistance ≤ 130 ℃ (ceramic liquid junction+high temperature gel)
High pressure type: withstand ≤ 10 bar (thickened glass film and strengthened sealing)
3、 Selection decision matrix
| operating condition characteristics | Recommended electrode types | core technology |
| pH 0-2, Containing trace amounts of HF | Fluoride ion resistant electrode | Specially made glass film, resistant to HF corrosion |
| pH 12-14, High temperature (≤ 80 ℃) | High lithium glass electrode | Anti alkali corrosion formula, high temperature stable type |
| Containing sulfides/heavy metal ions | Double salt bridge reference electrode | Special electrolyte for external salt bridge, resistant to poisoning |
| High viscosity/containing particulate matter | Open liquid network electrode | Anti clogging design for easy flushing |
| Hygiene grade requirements (CIP/SIP) | Sanitary PEEK/316L electrode | Complete structure, resistant to high temperature sterilization |
4、 Accurate matching of application scenarios
Electroplating industry: choose fluoride ion resistance+open liquid network electrode to deal with chromic acid
Chlor alkali industry: using high lithium glass and titanium alloy shell, resistant to concentrated alkali and chlorine gas environment
Chemical reaction kettle: adopting high-pressure type+double salt bridge design to adapt to pressure fluctuations and complex media
Wastewater treatment: equipped with anti pollution and PEEK material electrodes to cope with sludge erosion and chemical addition
5、 Maintenance and lifespan optimization
Proper selection can extend the electrode life to 18-24 months (6-12 months under normal operating conditions):
Calibration frequency: High acidity and alkalinity environments require weekly calibration, using pH 1.68/4.01/6.86/9.18/12.45 multi-point calibration
Cleaning plan: Soak strong alkali pollution in 5% dilute hydrochloric acid, and treat organic pollution with protease solution
Storage requirements: Long term disuse and soaking in 3mol/L KCl to avoid dehydration of the glass film
The selection of high acid-base pH electrodes is a technical decision rather than a simple procurement. Accurately matching the working conditions and electrode characteristics can avoid the dilemma of "calibration every three days and replacement every three months". Investing in professional grade electrodes not only yields precise data, but also long-term stable operational benefits and significantly reduced overall costs.