E-mail
15765947@qq.com
-
Phone
18011564711
-
Address
No. 248 Jinfu Road, Jinniu District, Chengdu
Chengdu Huaheng Instrument Co., Ltd
15765947@qq.com
18011564711
No. 248 Jinfu Road, Jinniu District, Chengdu
Model 905VProduct Introduction of Headspace Gas Analyzer for Pharmaceutical Bottles:
Pharmaceutical preparations packaged in small bottles often use inflatable packaging with a mixture of oxygen, nitrogen, or other gases added. Depending on the product, the capacity range of these small bottles is from 1 milliliter to 250 milliliters or larger. The analysis of oxygen content in small bottles presents some difficulties as the available sample size is usually quite small. For example, in a 1 milliliter sample bottle, there may only be 0.5 milliliters of top air. To directly analyze with an oxygen analyzer, the instrument must be able to analyze very small sample sizes.
Model 905VFunctional characteristics of headspace gas analyzer for medicine bottles:
Design of sensors with low sample size
905VThe sensor model is specifically designed for this purpose. The internal cavity of the oxygen sensor is only 0.1 milliliters, which is very small in volume. The sensor is located behind the front panel to avoid the consumption of sample volume. The sensor has input/output ports and a diameter of 0.03 inches. The sensor outlet is connected to a short narrow aperture tube for easy sample gas discharge and entry into zero or calibration gas.
Simplify calibration
The oxygen sensor output is very linear throughout the entire measurement range. This allows for single point calibration within the measurement range at any location. The span calibration is completed by testing the indoor air, and the reading should be 20.9% oxygen content. If necessary for internal testing programs, standard calibration can also be used.
Suitable for laboratories or production lines
Model 905VIt is a durable desktop instrument with a heavy metal casing, especially suitable for continuous use in routine packaging testing in laboratories or packaging lines. In addition to AC power, there is a battery for power operation (optional), allowing the instrument to be used on-site in remote locations without AC power.
Reliable design
Microprocessor based electronic products are used for sensor signal processing, battery charging, and on/off control to provide height and reliability. Durable, proprietary oxygen sensor, ensuring years of use, internally sealed, maintenance free. The output data of the sensor is simultaneously compensated for changes in environmental temperature.
Model 905VProduct features of headspace gas analyzer for pharmaceutical bottles:
.Analyze oxygen levels in all types of bottles
Battery powered (optional)
Range: 0.01 to 100% oxygen
Mobile, low capacity sampling probe
Economical and affordable
0-5 VDC analog output (optional)
Model 905VTechnical specifications for headspace gas analyzer for pharmaceutical bottles
Range: 0.0?100%Oxygen
Resolution: 0.01% oxygen
Degree:+/-1% reading; +0.1% below 10.0%
Zui low detection limit: 0.01%
Drift: less than 1% per month
Sensor: Electrochemical battery, expected lifespan of 4-5 years
Sampling method: water displacement method; Interactive method: Introducing gas samples for instrument analysis
Zui small sample requirement: 0.5CC
Sampling device: probe connected to side port needle and PTFE tube
Calibration: calibration with indoor air; Back thickness SPAN adjustment; Upper side fine spacing adjustment; Zero back adjustment
Calibration frequency: least once a year
Power supply: 100-240V (50-60Hz) charger includes a plug adapter that can be interchanged with the US or US (application required)
Backup power supply: optional 12V internal battery pack
Input voltage: 100-240V, 50-60Hz
Charging * -14 hours: (optional) Battery life
Size: 9.8in x 4 inches x 10 inches (249 x 102 x 254 millimeters)
Weight: 8 pounds (3.6 kilograms)
Standards: CE, ROHS
Including consumables: low volume sampling probe, zero volume puncture needle, tubing and tee connector, two plastic 1CC and 5ml syringes, 115/240Vac to 12Vdc power adapter
Including documents: user manual (including detailed calibration instructions, calibration certificate, and operating instructions for precautions), NIST calibration certificate
Product origin: Our products are manufactured in the United States
Next Article: Laser trace uranium analyzer