Magnetic heating stirrer is a commonly used equipment in laboratories, which integrates stirring and heating functions. However, long-term use is prone to various malfunctions. The following are common problems, causes, and targeted solutions:
1、 Abnormal mixing function
Phenomenon 1: Magnet does not rotate or the speed is unstable
Reason: External magnetic field interference (such as nearby strong magnets), uneven bottom of the container causing magnetic particle jamming, or motor aging.
Solution: ① Remove surrounding magnetic objects; ② Replace the container with a flat bottom surface; ③ If there is any abnormal noise from the motor, it needs to be returned to the factory for repair.
Phenomenon 2: Insufficient mixing force
Reason: Severe wear of the magnet, high viscosity of the solution, or overloading of the sample volume.
Solution: ① Replace with a new magnet; ② Use larger size magnets or adjust the speed; ③ Batch processing of high viscosity samples.
Phenomenon 3: Normal idling but stopping after loading
Reason: Insufficient motor torque or capacitor decay.
Solution: Contact the manufacturer to repair the motor and capacitor components.
2、 Heating function failure
Phenomenon 1: No heating
Reason: The heating wire is burnt out, the solid-state relay is damaged, or the power cord is disconnected.
Solution: ① Use a multimeter to check the continuity of the electric heating wire; ② Replace the relay with the same model; ③ Check the power cord interface.
Phenomenon 2: Slow heating or inability to reach the set temperature
Reason: Scale buildup on the heating plate affects thermal conductivity efficiency, sensor calibration deviation, or rapid heat dissipation.
Solution: ① Soak and clean the heating plate with oxalic acid; ② Re calibrate the temperature sensor; ③ Install insulation cover to reduce heat loss.
Phenomenon 3: Actual temperature far exceeds the set value
Reason: Temperature controller malfunction or heating tube out of control.
Solution: ① Immediately power off and replace the temperature control module; ② If the heating tube is short circuited, it needs to be replaced as a whole.
3、 Control system disorder
Phenomenon 1: Incomplete display on the digital display or malfunctioning buttons
Reason: The humid environment can cause the circuit to become damp, the buttons to have poor contact, or the motherboard program to malfunction.
Solution: ① Natural drying in a dry environment; ② Open the panel and clean the contact oxide layer; ③ Reset factory settings or upgrade firmware.
Phenomenon 2: Timing function failure
Reason: Aging of crystal oscillator components or memory overflow.
Solution: ① Replace the same frequency crystal oscillator; ② Close unnecessary programs to free up memory.
4、 Other mechanical issues
Phenomenon 1: Intense vibration or loud noise during operation
Reason: rotor eccentricity, lack of oil in bearings, or loose tabletop.
Solution: ① Correct rotor balance; ② Add high-temperature lubricating oil; ③ Tighten the table top fixing screws.
Phenomenon 2: The shell is hot and has a noticeable odor
Reason: Blockage of the fan leads to poor heat dissipation or short circuit of wires.
Solution: ① Clean the dust on the fan; ② Immediately stop the machine to check the circuit and replace the fuse if necessary.
5、 Daily maintenance suggestions
1. Regular maintenance: Clean the heating plate and magnet every month, and check the insulation layer of the wires every quarter;
2. Standardized operation: Avoid splashing corrosive liquids into the interior of the instrument and prohibit dry burning;
3. Environmental management: Place in a ventilated and dry place, away from high-frequency equipment to reduce interference.
The above methods can quickly locate and repair most faults. If it involves damage to core components, it is recommended to contact professional after-sales service. Reasonable maintenance can significantly extend the service life of equipment and ensure the safe and efficient conduct of experiments.