Operation Guide for Low Temperature Brittle Impact Testing Machine
Basic operating steps
preparation
According to《Preparation of Test Plan for GB5470-2008 Low Temperature Brittle Test Method
Inject industrial alcohol as a heat transfer medium into the cold bath, and the liquid level should be far from the cold slotBelow 20mm, it is advisable to use immersion fixtures
Check if the power wiring of the equipment is correct and ensure good grounding
Sample placement
According to the testing standards, place the sample in a low-temperature environment under the specified conditions
Rubber sample: elongation230mm (multi sample method 8 ± 0.3mm or single sample method 11 ± 0.5mm)
Plastic sample:3.6±0.1mm
temperature control
Set the target temperature (usually-Range of 60 ℃~0 ℃
Wait for the device to reach the set temperature and maintain a constant temperature (after constant temperature)Fluctuations within 3 minutes<± 0.5 ℃)
impact test
withApply impact load at a speed of 2m/s ± 0.2m/s
Record the damage of the sample
Count a sufficient number of samples (recommended not less than)10 of them)
data analysis
Calculate the brittleness temperature using statistical theory(Temperature at which 50% of the sample is damaged
Generate test report
Operation precautions
Safety Specifications
Before modifying settings and exporting signals during instrument operation, safety should be fully considered
Avoid using organic solvents such as alcohol and gasoline to wipe the instrument. It is recommended to use a dry cloth
Without permission, it is not allowed to disassemble or modify equipment to avoid abnormal movements or dangers
Equipment maintenance
Pay attention to preventing dust, wire ends, and other foreign objects from entering during installation to avoid causing malfunctions
During the experimental process, it is necessary to record data in a timely manner to ensure accuracy
Regularly check the accuracy of impact speed and temperature control
test environment
Ensure good ventilation in the laboratory
Avoid using equipment in damp or corrosive environments
Maintain sufficient operating space around the device
Common problem handling
Unstable temperature
Check if the liquid level in the cold bath is sufficient
Verify if the temperature sensor is working properly
Confirm that the equipment has good sealing performance
Impact velocity deviation
Calibrate the impact device
Check if the transmission components are worn out
Data anomaly
Recheck the placement of the sample
Verify whether the ambient temperature affects the test
Check the grounding condition of the equipment
standard reference
The following standard requirements should be followed during operation:
GB/T 5470-2008 "Determination of Brittle Temperature of Plastics by Impact Method"
ISO 812 International Standard
ASTM D2137 American Society for Testing and Materials standard
By following the detailed operation guide above, you can safely and effectively use the low-temperature brittle impact testing machine for material performance testing. For more specific operational guidance, it is recommended to refer to the user manual provided with the equipment or contact the equipment manufacturer for technical support.