1nitrogen-filled oven Technical specifications:
1Model:SC/DGG-9626AN
2Studio size(W*D*H,mm):800*600*1300
3Overall dimensions (W*D*H,mm):1035*825*1750(Actual appearance measurement shall prevail)
4. Temperature range:RT+10℃~250℃ 5. Constant temperature fluctuation: ±1℃
6. Temperature resolution: 0.1℃ 7 Heating rate:1.0~3.0℃/min
8. Temperature timing range:1-9999min 9. Power supply:AC380V/50Hz,4200W
10. Sample rack shelf5layer(Activity based)
IInitrogen-filled ovenMaterial for box production:
1The design of the test chamber is reasonable, and the box body is processed and formed by CNC machine tools, making it easy to operate.
2The inner liner is made of high-quality stainless steel plate(201 # or 304 # according to the contract,delivery perioddefault201 # stainless steel inner liner5One working day;定制304 # stainless steel inner liner 15One working day.)The shell isA3The board is treated with spray coating, making it smoother and more aesthetically pleasing.
3The insulation system uses ultra-fine glass fiber to fill the insulation area, and non-metallic high-temperature resistant materials are used at the connection between the inner and outer bile ducts, effectively reducing temperature conduction; The sealing of the box door is refinedhigh temperature resistanceSilicone rubber, so there is no aging or hardening phenomenon under high temperature conditions, and the sealing is more reliable;
4、 The air supply circulation system adopts low-noise, long-life, air conditioning type special fans, with multi wing centrifugal blades as the rotor.
IIItemperatureControl system:
1. The temperature controller adopts touch buttons and digital displayLEDdisplayPIDIntelligent control instrument;
2. Accuracy:0.1℃(display range); 3. Resolution: ±0.1℃;
4. Temperature sensor:PT100Platinum resistance tester; 5. Control method: Thermal balance temperature regulation method.
4、 Insulation system:
The insulation system uses ultra-fine glass fiber to fill the insulation area, and non-metallic high-temperature resistant materials are used at the connection between the inner and outer bile ducts, effectively reducing temperature conduction.
5Nitrogen filling system:
1. Pressure regulating valve: controls the nitrogen pressure, and under normal use, adjusts the nitrogen pressure to1~About 2 kilograms;
2. Flow meter: controls the size of nitrogen entering the box, with a range of 2L/min-10Lmin;
3. The nitrogen pipeline enters directly behind the box and is equipped with a 2-meter long 10mm nitrogen hose externally;
4. Nitrogen needscustomer-provided.
VIReference standard:
JB/T 5520-1991《Technical conditions for drying oven
How to translate7
Environmental requirements for use:1temperature5~40℃2
relative humidity≤85%RH3Power supply voltageAC380V±10% 50H
z 4、There is no strong vibration or corrosive gas influence around
eighty percentSafety protection device:Internal temperature over temperature protection: When the temperature inside the box exceeds the temperature set by the over temperature protector, the machine will automatically cut off the power supply to the heater and enter a balanced state for cooling
;Grounding protection; Fan overload protection.
nine |
Packing list: |
serial number |
category |
name |
unit quantity |
1 |
prepare |
note |
document |
1 |
|
2 |
Instruction Manual |
copy |
document |
1 |
|
3 |
Packing List |
copy |
document |
1 |
|
4 |
Certificate of Conformity |
copy |
spare parts |
2 |
Fuse wireonlyten、
warranty period:Non human caused faults, free warranty
twoYear.