Advantages of high-performance capillary electrophoresis: $n automatic injection system; $n comes with a built-in LCD display screen; $n comes with a built-in workstation and Windows 7 operating system, eliminating the need for an external computer; The built-in Wi Fi enables remote operation; $n automatic sampler; $n10 sample tray; The inlet end of the capillary tube can freely contact the sample bottle; $n capillary tube sleeve; $n High speed forced liquid cooling; $n Power Requirements: Voltage 100-240V, High Power 300W; $n Raw Data Channels: Detector Signal and Spectrum/Voltage/Current/Leakage Current/Power/Card Sleeve Temperature/Pressure/Lamp Voltage and Detector Temperature.
1 UC7010high performance capillary electrophoresisAdvantages:
1. Automatic sampling system;
2. Comes with a built-in LCD display screen;
3. Built in workstation and Win7 operating system, no need for external computer;
4. Built in Wi Fi enables remote operation;
5. 10 sample trays;
6. The capillary inlet end can be freely contacted, so the sample bottle.
II UC7010high performance capillary electrophoresisIntroduction to the structure:
(1) Sample injection/cleaning system:
1. Electromigration injection: injection voltage can be set arbitrarily from 0-30kV
2. Pressure injection: Injection pressure can be selected from 0-100kPa
3. Flushing method: Positive pressure flushing
4. Pressure range: 0-100kPa
5. Capillary tube: Quartz capillary tube, length ranging from 25-1000cm, inner diameter ranging from 25-150 μ m
6. Power requirement: 220V 50/60Hz
(2) UV detector
1. Wavelength range: 190-740nm
2. Wavelength accuracy:<± 1nm
3. Detection limit:<1 × 10-6g/ml
4. Baseline noise: ± 5 × 10-5AU at 254nm, t=1s
5. Baseline drift: 1 × 10-3AU/hour at 254nm, t=1s
(3) High voltage power supply
1. Output voltage: 0-30kV
2. Output current: 0-300 μ A
3. Power regulation rate: Input voltage variation ± 10%, output voltage variation 0.01%
4. Load regulation: 10V/100 μ A
5. Ripple: 0.1% of output voltage
6. Normal working conditions for high-voltage power supply: ambient temperature: 5-40 ℃
Relative humidity: ≤ 80%
7. Power supply voltage: single-phase AC 220V, allowable deviation ± 10%
8. Power frequency: 50Hz, allowable deviation ± 5%
(4) Pressure sensor
1. Output pressure: 0-100kPa
2. Allowable deviation: ± 2.5kPa
(5) Temperature sensor
1. Controllable temperature 15-80 ℃
2. Allowable deviation: ± 2 ℃
(6) Software
1. UC7010 is equipped with a chromatography workstation.
2. UC7010 is equipped with a built-in keyboard and wireless mouse, and a touch screen with a built-in computer. You can set a save directory to automatically save the electrophoresis map.