E-mail
86@511718.com
-
Phone
13261731718
-
Address
Celebrity Future Building
Beijing Huayi Tongtai Environmental Protection Technology Co., Ltd
86@511718.com
13261731718
Celebrity Future Building
YSI Pro Quatro portable multi parameter water quality analyzer from the United StatesProduct Introduction:
YSI ProQuatro handheld multi parameter instrument can measure dissolved oxygen, conductivity, specific conductivity, salinity, resistivity, total dissolved solids (TDS), pH, ORP, pH/ORP combination, ammonium (ammonia), nitrate, chloride and temperature in various combinations. It is also a very powerful laboratory instrument for BOD, pH, and ORP.
The EPA has approved the drinking water and wastewater compliance report.
Application areas of YSI ProQuatro portable multi parameter water quality analyzer in the United States:
Surface water and underground drinking water measurement
Measurement of agricultural water quality parameters
Non point source pollution in inland and coastal areas
Overflow from sewage treatment plants
Groundwater near sewage treatment plants that may be contaminated
Overflow caused by rainstorm
Water for entertainment projects
地表水工程
Wetland monitoring
Research on Salt Tide Invasion
Laboratory BOD testing
The food and beverage industry, mainly consisting of wine and beer
Comparison of data as a long-term unmanned monitoring device
University research project
Configuration List of YSI ProQuatro Portable Multi Parameter Water Quality Analyzer in the United States:
YSI ProQuatro portable multi parameter water quality analyzer
ProQuatro Multi Parameter Water Quality Analyzer Host
4-meter cable, four probe interface 10102030-4
Dissolved oxygen electricity * YSI2003
PH electricity * YSI1001
ORP Electric * YSI1002
Electrical conductivity * YSI5560
Ammonia ion selective charge * YSI1004
Chloride Electricity * YSIYSI1005
Nitrate Electricity * YSI1006
Membrane kit YSI5908
PH buffer YSI3821, YSI3822, YSI3823
Conductivity calibration solutions YSI3167, YSI3168, YSI3169
Oxidation reduction ORP standard solutions YSI3682, YSI3683, YSI3684
YSI Pro Quatro portable multi parameter water quality analyzer from the United StatesTechnical Specifications:
Technical parameters of YSI ProQuatro portable multi parameter water quality analyzer (4-meter cable+probe):
parameter |
Measuring principle |
measurement range |
resolution |
accuracy |
calibration |
working depth |
|
dissolved oxygen |
%Air saturation |
*Spectral method or primary battery method |
0-500% |
0.1% or 1% air saturation (optional) |
0-200%: ± 2% of the reading or 2% air saturation, whichever is greater; 200-500%: ± 6% of reading |
1 point or two-point calibration with 0% |
|
Milligrams per liter |
0-50mg/L |
0.1 or 0.01mg/L (Optional) |
0-20mg/L: ± 2% of the reading or 0.2mg/L, whichever is greater; 20-50mg/L: ± 6% of reading |
1 point or two-point calibration with 0% |
|||
temperature |
Field durable cable |
-5 to 70 ℃ |
0.1 |
±0.2℃ |
4m |
||
Laboratory grade cables |
5 to 70 ℃ |
0.1 |
±0.35℃ |
||||
Conductivity* |
Four electric * circulating type Conductivity measurement tube method |
0-200mS/cm |
0.001-0.1mS/cm (Depending on the range) |
± 0.5% of the reading or 0.001mS/cm, whichever is greater (4m cable); ± 1% of the reading or 0.1ppt, whichever is greater (20m cable) |
one o'clock |
||
salinity |
Calculated from conductivity and temperature |
0-70ppt |
0.01ppt |
± 1% of the reading or 0.1 ppt, whichever is greater 0.2 |
one o'clock |
||
pH |
Glass composite electrochemical method |
0-14 |
0.01 |
±0.2 |
1, 2, 3, 4, 5 or 6 o'clock (Optional) |
||
ORP |
Platinum electrochemical method |
-1999 to 1999mV |
0.1mV |
±0.2mV |
one o'clock |
||
Ammonium nitrogen |
Ion selective electrochemical method |
0-200mg/L-N(0-40) |
0.1mg/L-N |
The reading is ± 10% 2mg/L-N, whichever is greater |
1, 2 or 3 o'clock (Optional) |
17m |
|
Nitrate nitrogen** |
Ion selective electrochemical method |
0-200mg/L-N(0-40) |
0.1mg/L-N |
± 10% of the reading or 2mg/L-N, whichever is greater |
1, 2 or 3 o'clock (Optional) |
17m |
|
Chloride** |
Ion selective electrochemical method |
0-1000mg/L(0-40) |
0.1mg/L |
± 15% of the reading or 5mg/L, whichever is greater |
1, 2 or 3 o'clock (Optional) |
17m |
|
|
Total dissolution solid |
Calculated from conductivity and temperature |
0-100g/L (can input 0.30 to 1.00, default value is 0.64) |
0.001, 0.01, 0.1 g/L |
0.4kPa (temperature variation within ± 15 ℃ of calibration point) |
|||
atmospheric pressure |
Piezoresistive method |
50-110kPa |
0.01kPa |
one o'clock |
|||
|
Note: *The parameters calculated from this include resistivity, salinity, specific conductivity, and total dissolved solids **Ionic electricity * only applicable to freshwater ***Additionally: US, NIST, or custom buffer solutions | |||||||
Host technical specifications:
parameter |
measurement range |
resolution |
accuracy |
pH |
-2.60 to 16.60 |
0.1mV (0.01pH unit) |
± 0.1mV (± 0.01pH unit) |
Oxidation-reduction potential |
-1999 to+1999mV |
0.01mV |
±0.5mV |
conductivity |
0.0-200 mS/cm |
0.001mS/cm or 0.1 μ S/cm-0.1mS/cm (depending on the range) |
± 0.3% ± 1 digit of the full range |
dissolved oxygen |
0.00-90mg/L 0-550% |
0.01mg/L, 0.1% |
± 0.2% of full range (550% air saturation) Previous position (at 10 o'clock, using 1.25PE cover film) |
temperature |
-10 to 100 |
0.1℃ |
Full range ± 0.1% 1-digit |