Weihai Laboratory Electric Heating Stainless Steel Reactorparameter
Design pressure: ≤ 12.5Mpa Working pressure: ≤ 9.8MPa
Design temperature: ≤ 350 ℃ Operating temperature: ≤ 300 ℃
Heating power: 1.5-7.5KW
Power supply: 220V, 380V.
Mixing speed: 50-750r/min, adjustable motor power: 55-550W
Main materials: 304, 316L, 321, C276, TA, Ni, etc
Sealing form: Magnetic seal.
Sealing gasket: stainless steel wrapped gasket, graphite composite gasket, octagonal gasket, PTFE gasket (working temperature below 180 ℃)
Mixing methods: magnetic stirring, mechanical stirring.
Mixing slurry forms: propulsion type, slurry type, anchor type, six blade turbine, Bourmakin type optional
Laboratory magnetic stirring high-pressure reactorController: The controller is equipped with an intelligent digital temperature control meter with self-tuning function, which has automatic temperature control function; The panel displays the speed, temperature inside the kettle, heating voltage, motor current, working time, etc. The speed is adjustable.
When conducting reactions with different media in a reactor, it is necessary to first investigate whether the media corrode the main material. Special orders are required for chemical reactions that cause rapid reactions, generate large amounts of gas, or are flammable and explosive at high temperatures, as well as reactions that cause severe corrosion to stainless steel due to ultra-high pressure, ultra-high temperature, or chloride ions, fluoride ions, etc. in the medium!
2When loading the reaction medium, it should not exceed the laboratory reactor body2/3liquid level!
3It is strictly prohibited to disassemble under pressure during or after work!It is strictly prohibited for the kettle to work under overpressure and overheating conditions!
4When cleaning the high-pressure reactor, special attention should be paid not to let water or other liquids flow into the heating furnace to prevent the heating furnace wire from burning out. If heating the jacket heat transfer oil, be careful not to mix water or other liquids into it when adding heat transfer oil. The oil level of the heat transfer oil should be checked periodically!
5If there is abnormal noise inside the isolation sleeve during operation, the machine should be stopped to release pressure and the mixing system should be checked for any abnormal conditions. Regularly check the swing amount of the mixing shaft. If the swing amount is too large, the bearing or sliding shaft sleeve should be replaced in a timely manner!
6Connect the corresponding voltage according to the operating voltage of the controller, single-phase220V, three-phase380VDo not connect the live and neutral wires of the controller in reverse;Wiring work should not be carried out during power transmission to prevent electric shock!To ensure the normal operation of the controller and the personal safety of the staff in Weihai, it is necessary to connect the grounding wire!The controller itself is not explosion-proof, so it should avoid flammable and explosive environments such as oil and gas!
7After the final reaction temperature is determined, it is not allowed to change the temperature setting value during the heating process to avoid significant temperature overshoot!
8Under the premise of connecting the power supply, even if the heating and stirring switches are not turned on, the output terminals on the back panel of the controller are still live, and it is not allowed to pull or touch any plugs during operation!
9The working time display is an internal microcontroller8031By accumulating internal frequency division for timing, each power on/off will cause the internal accumulated data to disappear. During operation, it should be noted that the working time is displayed in minutes, and the maximum working time is9999part(166.5hour)!
10WeihaiLaboratory Electric Heating Stainless Steel ReactorThe correct detection position of the speed display is crucial for the stable display of the speed. This position has been adjusted at the factory and should not be adjusted arbitrarily!
11Every time the machine is turned on, it is required that any button should be in its initial state. This requires the operator to develop good working habits and turn the knob back to the minimum position after each operation to prevent excessive current from causing significant damage to the controller during the next power on!
12Regularly inspect various instruments and blasting relief devices to ensure their accurate and reliable operation. The working environment of the equipment should comply with safety technical specifications!
13When installing the laboratory reaction kettle, connect the explosion relief port to the outside through a pipeline!