GermanyIntroduction to WTW OxiTop Control 6/12 Laboratory BOD Analyzer
biochemical oxygen demandThe determination of BOD is one of the important parameters in water resource management. BOD is not only an important indicator for evaluating water quality and wastewater treatment effectiveness, but also an important basis for planning and designing wastewater treatment equipment.
BreathThe pressure gauge method is a simple and reliable way to measure BOD. OxiTop is a BOD measurement system based on this principle. Microorganisms in water samples consume oxygen, and the CO2 produced is absorbed by NaOH, forming negative pressure in a closed measuring bottle. OxiTop directly reads the mg/L BOD measurement value by measuring the pressure change.
Advantages and Characteristics
◆ Breath(Pressure gauge method) measuring BOD
◆ OxiTop Control 6 can simultaneously measure many 6 samples
◆ OxiTop Control 12 can simultaneously measure many 12 samples
◆ By using different sample sizes, a large measurement range can be achieved4000mg/L
◆ The OxiTop-C measuring head is operated by the OC100 controller through an infrared interface
◆ The OxiTop-C measuring head has an automatic temperature function, which automatically delays the measurement for 1-3 hours when the sample has not reached a constant temperature, until the temperature stabilizes
◆ The OxiTop OC 100 controller can start, manage, store, and track 100 measurement heads simultaneously or in groups. It features a large screen display and graphical display of data for each measurement head
◆ choose and buyPC communication cables and software can download measurement data to PC for comprehensive analysis and management
◆ Long battery life, approximatelyReplace every 2 years
Product parameters
|
OxiTop-C sensor head |
|
|
measuring principle |
Breathing method (pressure gauge method) |
|
Measurement parameters |
BODn |
|
Pressure Range |
500 ~ 1350hPa(mbar) |
|
accuracy |
± 1 measurement value |
|
resolution |
1hPa |
|
display |
LED indicator light |
|
Working power supply |
Lithium battery (260mAh), 2 CR2430 (3V) |
|
BATTERY LIFE |
About 2 years |
|
Storage temperature |
-25℃~ +65℃ |
|
operation temperature |
+5℃~ +50℃ |
|
Dimension |
70mm x 70mm diameter, approximately 95g |
|
OxiTop OC 100 Controller |
|
|
measuring range |
0 ~40/80/200/400/800/2000/4000mg/L BOD |
|
Measurement data group |
180-360 per measurement cycle |
|
Measurement period |
0.5 hours~99 days |
|
display |
LCD graphic display, 64 × 128 dot matrix |
|
data interface |
Infrared (with measuring head), RS232 connected to PC |
|
Working power supply |
3AA × 1.5V alkaline battery with a lifespan of 1000 measurements |
|
storage temperature |
-25 ~ +65℃ |
|
operation temperature |
+5 ~ +40℃ |
|
Dimension |
45×100×200mm,390g |
|
A magnetic stirrer |
|
|
Working power supply |
100-240VAC adapter, output 18VAC/1A |
|
Power consumption |
Up to 24VA |
|
Blender speed |
Program control, 180~450 revolutions per minute |
|
Dimension |
IS 6:265×181×67mm,1.6kg; IS 12:350×266×67mm, 3.2kg |
Standard configuration
|
Item number |
configuration |
|
208201 |
OxiTop Control 6, 6 BOD measurement units, including OC100 controller, 6 OxiTop-C measuring heads, 6 brown 510ml measuring bottles and rubber cylinders, 6 stirring rods and 1 rod, 1 IS6 stirrer, 100-240VAC adapter, 50g CO2 absorbent, 25ml nitrification inhibitor, 1 164/432ml overflow bottle, 1 record paper, user manual |
|
208204 |
OxiTop Control 12, 12 BOD measurement units, including OC100 controller, 12 OxiTop-C measuring heads, 12 brown 510ml measuring bottles and rubber cylinders, 12 stirring rods and 1 rod, 1 IS12 stirrer, 100-240VAC adapter, 50g CO2 absorbent, 25ml nitrification inhibitor, 1 164/432ml overflow bottle, 1 record paper, user manual |
optional accessories
|
902842 |
AK 540/B, Infrared connection cable, 6-pin plug to RS-232 interface, including 9/25 adapter |
|
208990 |
Achat OC, PC communication software for OxiTop OC100 or OC110, measurement data processing and file archiving |
|
Negotiation |
Measuring head, measuring bottle, marking ring, stirring rod, absorbent, nitrification inhibitor, overflow bottle, constant temperature incubator, etc |