Product Application
Usually used in conjunction with display instruments, recording instruments, electronic computers, etc. Directly measure the temperature of liquid, vapor, gas media, and solid surfaces within the range of 0 ℃ to 1300 ℃ in various production processes.
Product Features
Less thermal response time, reducing dynamic errors; Flexible installation and use; Large measurement range; High mechanical strength and good pressure resistance.
Product execution standards
IEC584 IEC1515 GB/T16839-1997 JB/T5582-9
Working principleThe electrodes of armored thermocouples are composed of two conductors of different materials. When there is a temperature difference between the measuring end and the reference end, a thermoelectric potential is generated, and the working instrument displays the temperature value corresponding to the thermoelectric potential.
Measurement range and tolerance(t is the actual measured temperature value)
|
model
|
Graduation
|
Tolerance levelⅠ
|
Tolerance levelⅡ
|
|
Allowable deviation
|
Measuring range℃
|
Allowable deviation
|
Measuring range℃
|
|
WRNK
|
K
|
±1.5℃
|
-40~+375
|
±2.5℃
|
-40~+333
|
|
0.004│t│
|
375~1000
|
±0.0075│t│
|
333~1200
|
|
WRMK
|
N
|
±1.5℃
|
-40~+375
|
±2.5℃
|
-40~+333
|
|
±0.004│t│
|
375~1000
|
±0.0075│t│
|
333~1200
|
|
WREK
|
E
|
±1.5℃
|
-40~+375
|
±1.5℃
|
-40~+333
|
|
±0.004│t│
|
375~800
|
±0.004│t│
|
333~900
|
|
WRFK
|
J
|
±1.5℃
|
-40~+375
|
±1.5℃
|
-40~+333
|
|
±0.004│t│
|
375~750
|
±0.004│t│
|
333~750
|
|
WRCK
|
T
|
±1.5℃
|
-40~+125
|
±1.5℃
|
-40~+133
|
|
±0.004│t│
|
125~350
|
±0.004│t│
|
133~1000
|
|
WRPK
|
S
|
±1℃
|
0~1100
|
±2.5℃
|
0~600
|
|
|
1100~1600
|
±0.0025│t│
|
600~1600
|
thermal response time
|
Pipe sleeve diameter
|
Φ3
|
Φ4
|
Φ5
|
Φ6
|
Φ7
|
other
|
|
Thermal response time (S)
|
≤3
|
≤4
|
≤8
|
≤12
|
≤15
|
Installation method