Visual dust explosion propagation/blockage small-scale testing device

NegotiableUpdate on 05/06
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin

Overview

This device is based on the national standard GB/T16428 and integrates precision machining, wireless control technology, and computer technology. The dust storage tank is pressurized, automatic powder spraying, ignition data acquisition, wireless data transmission and storage, and data analysis to draw flame temperature/time curves are all automatically completed by the system control. It is a functional explosion testing equipment. $n computer wireless monitoring, data wireless transmission, automatic drawing of flame temperature/time curve, human-machine separation, ensuring personal safety.

Product Details

1Control method:

1.1Computer wireless monitoring:

Wireless transmission of control parameters and test data, human-machine separation, ensuring the personal safety of test personnel. The test device automatically sprays dust and ignites according to the set parameters, automatically collects flame temperature and stores it in a data file, and draws flame temperature/Time curve, data can be transferred toEXCELFurther analysis and processing will be conducted.

1.2 Manual control:

By simulating a keyboard, manual dust spraying and ignition, data collection and storage, container cleaning, and other operations can be performed, and data files can be read into a computer for analysis and processing.

2Structure and main configuration:

2.1Explosive container: square pipe, volume3.6 LMaximum work pressure:2MPaThe material isS30408Stainless steel)

2.2Pipeline length: total length1000mm

2.3The cross-section of the pipeline is square, with inner side length60mm

2.4Optical system: The front and rear sides of the pipeline are transparentHigh strength tempered glassVisual Area: Long900mm*wide60mm,The window surface is strictly parallel and pressure resistant2MPa.

2.5Test location: The left and right sides of the square pipeline areS30408Stainless steel, on the left side1/4height sum3/4Install one temperature sensor at each of the two height positions,on the right1/3height sum2/3Install explosion suppression system interfaces at two different heights.

2.6Ignition position: from the bottom100mmThere is a pair of ignition electrode mounting holes at the location.

2.7Safety protection: A large-diameter fire-resistant and pressure relief device is installed at the top of the pipeline, which has the function of fire-resistant and pressure relief.

2.8Dual channel configuration:

通道1Dust control system

Configure volume150mlThe dust storage tank and high-speed injection valve have a maximum working pressure of:1MPaThe dust pressure can be adjusted through a testing and analysis system.

通道2Explosion suppression system

Configure volume50mlThe explosion suppressant storage tank (powder)/Gas and high-speed injection valve, maximum working pressure:1MPaThe injection pressure can be adjusted through a testing and analysis system.

3Synchronous control and multi-channel data acquisition system:

configuration6Channel data acquisition system, set control parameters, the experimental device automatically sprays dust and ignites according to the set parameters, and automatically collects dataflame temperatureAnd store it in a data file. automatic drawingflame temperature/timeCurve, data can be transferred toEXCELFurther analysis and processing will be conducted.

3.1Flame temperature detection range:0~1200℃ Resolution:1℃ precision0.25%FS

configuration2Support high-speed flame temperature sensor.

3.2Data sampling interval:0.2msMaximum sampling time:6S;

4Test conditions:

4.1Initial pressure: local atmospheric pressure

4.2Initial temperature: room temperature

4.3Pressure of dust storage tank:0~1.0 MPa

4.4Pressure of explosion suppressant storage tank:0~1.0 MPa

5, Point Fire: Electric spark energy generator, remote wireless control ignition and detonation.

5.1Range of ignition energy adjustment:1~3000mj

5.2Adjustable ignition energy accuracy:1mj

5.3The ignition electrode is detachable

6External device driver: The external device synchronization triggering function module can synchronously start external high-speed cameras, shadowgraphs, and other external devices.

7Cooling method: natural cooling

8Workstation: Laptop (user provided)

9Electricity Source:220 V±10%AC 50Hz±2Hz Power: <2500W

10Working environment:

Ambient temperature10℃~40℃

ambient humidity:30%~80%RH