Phone
13824460726
-
Address
Room 901, No. 12 Huaxin 1st Street, Haizhu District, Guangzhou City
Guangzhou Kefu Instrument Co., Ltd
13824460726
Room 901, No. 12 Huaxin 1st Street, Haizhu District, Guangzhou City
Standard configuration:
Inner cavity: made of stainless steel material, 1.4404 (ASTM3316L), sealed and welded, with detachable brackets on the sides for easy cleaning, including heating partitions;
Internal: heated shelf, made of aluminum material 3.3547 (ASTM B209);
Box body: embossed stainless steel outer surface, galvanized steel plate back, glass stainless steel operation panel, with multiple display and input modules, explosion-proof glass doors inside and outside;
Plug: optional;
Installation: 4 supports;
Interface: USB, Optional Ethernet and printer interfaces.
Model and size specifications |
200 |
400 |
500 |
||
Stainless steel inner cavity |
volume |
rise |
29 |
49 |
101 |
width |
mm |
385 |
385 |
545 |
|
height |
mm |
305 |
385 |
465 |
|
depth |
mm |
250 |
330 |
400 |
|
Maximum number of shelves |
quantity |
3 |
4 |
4 |
|
Distance between shelves |
mm |
75 |
75 |
75 |
|
Maximum load capacity of each shelf |
kg |
20 |
20 |
20 |
|
Maximum load capacity of the box |
kg |
60 |
60 |
60 |
|
External embossed stainless steel |
width |
mm |
550 |
550 |
710 |
height |
mm |
600 |
680 |
760 |
|
Depth (without door handle, door handle depth 38mm) |
mm |
400 |
480 |
550 |
|
Safety glass door, embossed stainless steel frame, internal spring-loaded safety glass door, external splash screen ESG |
|
□ |
□ |
□ |
|
Sealing of the door |
Circular silicone grease sealing strip |
|
□ |
□ |
□ |
temperature |
Electronic microprocessor temperature controller with Pt100 and self diagnostic system |
|
□ |
□ |
□ |
Each heating shelf comes with a 4-wire Pt100 sensor |
|
□ |
□ |
□ |
|
temperature range |
°C |
5K above ambient temperature~ +200 |
|||
Set temperature range |
°C |
+20 ~ +200 |
|||
Temperature changes over time (aluminum heated shelf) |
K |
≤ ±0.3 |
|||
+160Temperature uniformity at ° C/50mbar(Surface) (Aluminum heated shelf) |
K |
≤ ±2 |
|||
vacuum degree |
Digital electronic pressure control (programmable, up to 40 stages, each stage can be individually adjusted) with a pressure range of 10mbar to1100mbar. It can control the intake, temperature, and vacuum degree programs, accelerate sample drying, and reduce moisture content |
|
□ |
□ |
□ |
When the door is opened, air is quickly drawn in without changing the selected vacuum setting point |
|
□ |
□ |
□ |
|
Allow minimum vacuum degree |
mbar |
0.01 |
|||
Maximum leakage rate |
bar/h |
0.01 |
|||
monitor |
The microprocessor temperature monitor is used for over temperature protection (protection level 3.1), and the Pt100 sensor contains a fault diagnosis device, with visual and audible alarms |
|
□ |
□ |
□ |
High and low temperature monitor |
|
□ |
□ |
□ |
|
Temperature monitoring strip, automatically linked to set point (ASF) |
|
□ |
□ |
□ |
|
Multi layer over temperature protection (MLOP) for each heating shelf |
|
□ |
□ |
□ |
|
Stop heating when a malfunction occurs |
|
□ |
□ |
□ |
|
Mechanical temperature limiter (TB) |
|
□ |
□ |
□ |
|
Sound alarm: high temperature and low temperature |
|
□ |
□ |
□ |
|
Timer function |
Real time/weekly program editing function (such as Monday to Friday) |
|
□ |
□ |
□ |
Timer displaying remaining running time: programmed through the control panel or memory card, up to 40 stages (1 minute per stage)-999h) unlimited number of stages through computer and free software programming |
|
□ |
□ |
□ |
|
file |
Memory 1024KB, with loop memory function, including set temperature, actual temperature, error message, time and date, can store up to 3 months of data, with a collection frequency of 1 minute |
|
□ |
□ |
□ |
“CeThe "lsius" software is used to control and record temperature and pressure |
|
□ |
□ |
□ |
|
set up |
Three point temperature calibration can be achieved on the control panel |
|
□ |
□ |
□ |
Language: D/UK/E/F/I |
|
□ |
□ |
□ |
|
connect |
DN16 flange is used for vacuum connection and air inlet |
|
□ |
□ |
□ |
Detailed data |
Electrical load at 230V, 50/60Hz(Load the maximum number of heating shelves) |
W |
1200 |
2000 |
2400 |
Standard attachments |
Detachable but internal bracket - made of stainless steel material (ASTM)316L)-Integrated horizontal guide rod with heating shelf placement |
|
□ |
□ |
□ |
Heating shelf interface |
quantity |
2 |
|||
Heating Shelf - Aluminum Material (ASTM)B209)-Equipped with integrated large-area heating, including local temperature sensing(4-wire Pt100 sensor)); Provide separate high-temperature protection for each shelf. |
quantity |
1 |
|||
Work Calibration Certificate (+160 ° C, 50mbar, at the center of the shelf), each shelf has a separate calibration certificate |
|
□ |
□ |
□ |
|
Packaging size/vacuum oven |
Net weight/gross weight (including cardboard box) |
kg |
58/64 |
82/90 |
120/134 |
Packaging dimensions width/height/depth |
cm |
67/81/54 |
67/89/63 |
82/97/67 |
|
Packaging size/pump module |
Net weight without/with pump |
Kg |
26/40 |
30/45 |
41/56 |
Gross weight without/with pump |
kg |
32/46 |
38/53 |
57/69 |
|
Packaging dimensions width/height/depth |
cm |
67/70/54 |
67/78/63 |
82/97/67 |
|
Order model: Vacuum oven |
|
|
VO200 |
VO400 |
VO500 |