This laboratory ultrapure water system with * function, P series, can produce Type I ultrapure water from RO water, distilled water, or deionized water. The water quality of the product meets or exceeds the CLSI and CAP Class I water standards.
The laboratory for this feature *Ultra pure water system, P seriesType I ultrapure water can be produced from RO water, distilled water, or deionized water. The water quality of the product meets or exceeds the CLSI and CAP Class I water standards.
- Simple and intuitive way to fetch water through the water handle or center console
- The center console is an 8-inch waterproof touch screen, and users can control the pure water system through the center console.
- A single pure water host can be equipped with multiple water intake handles ("1+N mode")
- The water outlet speed of the handle can reach up to 2L/minute, and it can also take water quantitatively
- Consumables with RFID tags can track the production date, installation date, replacement period, excess water volume, and usage time throughout the entire process
- Both the controller and the center console can be wirelessly controlled (optional)
- Real time TOC and resistivity/conductivity monitoring
- Users can view basic information (water quality, system status, warning messages) at any time on the water intake handle and center console, with data storage up to 2 years and downloadable
The graphical interface of the monitor can guide users in system operations, troubleshooting, and maintenance work, and multiple permission levels can be set through passwords
Application scope: HPLC mobile phase preparation; Prepare reagent blank solution; Dilute samples using GC, HPLC, ICP-MS, AA, and other analytical techniques; Preparation of buffer and culture medium for mammalian cell culture; Preparation of molecular biology reagents.
Certifications:The VWR pure water system comes with a certificate of conformity to ensure that the product has been fully manufactured and tested according to standard operating procedures, and provides calibration certificates for the temperature and resistivity meters built into the system. Consumables will provide quality certificates along with the goods. The production factory has obtained ISO 9001 (2015) and ISO 140001 certifications.
Ordering NoticeA complete pure water system includes a pure water host and necessary accessories. Other optional accessories and consumables, including wall mounted mounting brackets, infinite level water tanks, foot switches, and wireless upgrade kits for water intake handles, can be ordered separately.
Ultra pure water system, P seriesspecification
| model |
P |
P |
P |
| water supply |
EDI water, RO water, distilled water, deionised water |
EDI water, RO water, distilled water, deionized water |
EDI water, RO water, distilled water, deionised water |
| Flow rate (l/min) |
Max. 2 |
Up to 2L/minute |
Up to 2L/minute |
| Total organic carbon content |
<5 ppb |
<5 ppb |
<5 ppb |
| +Electrical resistivity at 25 ° C |
18,2 MΩ·cm (Type I) |
18.2 MΩ·cm (Type I) |
18.2 MΩ·cm (Type I) |
| particle |
No particles (with a 0,2 μm final filter or Bio filter) |
No particles (using 0.2um terminal filter or Biofilter terminal filter) |
No particles (using 0.2um terminal filter or Biofilter terminal filter) |
| endotoxin |
<0,001 EU/ml (with Bio filter) |
<0.001 EU/ml (using Biofilter terminal filter) |
<0.001 EU/ml (using Biofilter terminal filter) |
| microorganism |
<0,1 cfu/ml (with a 0,2 μm final filter or Bio filter) |
<0.1 cfu/ml (using 0.2um terminal filter or Biofilter terminal filter) |
<0.1 cfu/ml (using 0.2um terminal filter or Biofilter terminal filter) |
| RNA enzyme |
<0,5 pg/ml (with Bio filter) |
<0.5 pg/ml (using Biofilter terminal filter) |
<0.5 pg/ml (using Biofilter terminal filter) |
| DNA enzyme |
<10 pg/ml (with Bio filter) |
<10 pg/ml (using Biofilter terminal filter) |
<10 pg/ml (using Biofilter terminal filter) |
| Width x Depth x Height (mm) |
300×480×510 |
300×480×510 |
300×480×510 |