UV visible spectrophotometer
1. Main performance characteristics
1.1 Dual beam measurement method, with five adjustable bandwidth levels;
*1.2 Real time dark current automatic correction technology;
*1.3 Instrument self inspection andGLP/GMPGuidance function;
*1.4 Wavelength driven subdivision, high-voltage synchronous servo technology:Achieve maximum scanning speed2000nm/min(movement speed3000nm/min);The minimum wavelength adjustment amount is0.01nm;
*1.5 Built in protein/The nucleic acid analysis function can be tested separatelydsDNAThessDNATheRNATheOligoWaiting for nucleic acid samples and utilizing direct ultraviolet absorption methodBradfordTheBCATheLowryTesting protein samples using equal methods;
*1.6 Trace sample measurement:Micro measurement: applicable sample size≥50uLAnd instrument noise≤0.004Abs(340nm,1min);
*1.7One computer can control simultaneously8setUV2600;
1.8 Capable of photometric measurement and spectroscopy/Time scanning, quantitative measurement, protein analysis/Nucleic acid measurement and other functions.
*1.9Integrated cast aluminum base ensures stability of optical system
2.Technical specifications:
2.1Wavelength range:190~900nm
2.2Spectral bandwidth:0.2,0.5,1,2,5nm(Five speed automatic adjustable)
2.3 Wavelength accuracy:±0.3 nm
2.4Wavelength repeatability: ±0.1 nm
2.5 Minimum wavelength adjustment amount:0.01nm
*2.6Range of photometric values:Abs:-2.000~3.000 %T:0~300%T Conc:0~9999 ;E(S),E(R):0~600
*2.7Accuracy of photometric values(NIST930DStandard color filter): ±0.3%T;±0.002Abs (0~0.5 Abs);±0.004Abs(0.5~1.0 Abs);±0.008Abs(1.0~2.0 Abs)
*2.8Repeatability of photometric values(NIST930DStandard color filter):±0.15%T;±0.001 Abs (0~0.5 Abs);±0.002 Abs(0.5~1.0 Abs);±0.004 Abs(1.0~2.0 Abs)
*2.9Wavelength scanning speed:1 nm/minThe5nm/minThe20 nm/minThe120 nm/minThe300 nm/minThe600 nm/minThe1000 nm/minThe2000nm/min
*2.10Scattered light: less than0.010%T(220nm Nal,340nmThe360nm NaNO2)
2.11Baseline stability:0.0004 Abs/h(500nm,2After preheating for hours)
*2.12Noise level:±0.0004 Abs (500nm)
2.13Baseline straightness:±0.001 Abs(200~850nm)
2.14Light source:WIThe lamp,D2light
2.15Light source switching: automatic switching,325~370nmUsers have the freedom to choose
2.16Detector: photomultiplier tube
2.17Overall dimensions:710(W)×630(D)×268(H)mm
Sample Room Size:124(W)×300(D)×142(H)mm
2.18Weight: Approximately50kg
2.19Power Supply:220VAC(50/60HZ);300VA
3.Software features
3.1 Measurement functions: spectral scanning, time scanning, quantitative analysis, multi wavelength measurement, kinetic testing, nucleic acid testing/Protein specific analysis kit;
*3.2 control function:Wavelength shift, automatic zero calibration, instrument wavelength automatic calibration, dark current deduction, automatic sampler and automatic six cell program control,Optical path correction, etc;
3.3 measurement:Startup initialization, setting of measurement conditions, standards orEXCELFormat report output or storage;
*3.4 The data shows:Simultaneously display spectra and data, and overlay spectra;
*3.5Quantitative method: 1-3Quantification of individual wavelengths, constant input method, standard curve method(1time.2time.3Quadratic functions and line fitting)QuantitativeABSCoefficient calculation;
*3.6Data processing: differentiation (differentiation), smoothing, operation(Graph and Graph,Graph and numerical values)Calculation of kinetic activity values, automatic calculation of mean, standard deviation, and relative standard deviation.
UV visible spectrophotometer