UK Ionic TVOC Fixed PID Gas Monitor
Intrinsically safe fixed PID gas monitoring device designed for continuous monitoring of volatile organic gases (VOCs)
TVOCSpecially designed for continuous monitoringVOC`sDesign and manufacturing are ideal safety equipment for typical volatile organic gas high-risk areas such as manufacturing and processing industries.
Features:
Adopting high-precision, moisture-proof and pollution proof grid electrode technology
Intrinsic safety applicable to hazardous areas—No need for additional explosion-proof barriers
Optional range:0-10ppm,0-100ppm,0-1000ppm
Resolution: 0.01ppm, 0.1ppm, 1ppm
Diffusion sampling—No need for an electric pump
4—20mAIntelligent analog output
5—28VWorking Voltage
-20℃—50℃Operating Temperature
LEDand4—20mAOutput fault indication
equip10.6eVfluorescent tube
Electronic digital display
easy to replacePID, easier to maintain
Easy calibration procedure
UK Ionic TVOC Fixed PID Gas Monitor
TVOCIt is a fixed photoionization(PID)Gas monitoring device, capable of continuously monitoring total volatile organic gases. equipped with4—20mAIntelligent simulation outputTVOCCan be easily integrated intoDCSControl system to achieve high concentration in the working environmentVOCSecurity alarm.
Compared to pump suction systems,TVOCUtilizing diffusion based samplingTechnology can reduce pollution in the testing room, thereby reducing the use of detection lampsCleaning frequency and maintenance work.TVOCEasy to install and maintainAnd calibration. At the same time, no thermal effect permit is required, andPIDsensingThe device can be replaced and maintained within seconds.
Application fields:
manufacturing
processing industry
Refineries and petrochemical enterprises
Offshore drilling monitoring
chemical plant
sewage treatment plant
pharmaceutical factory
Indoor Air QualityIAQmonitoring
Pulp and paper industry
Solvent extraction system
Industrial paint and electroplating industry
Technical Specifications
Sensors(PID):
adopt10.6eV Ion ScienceGrid electrode detection lamp
Intrinsic safety certification:
Admission protection level:
IP—65conduit
IP—45sensor
power supply
5—28VDC, * * * Big65mA
output:
4—20mA(requirement8—35VDCpower supply)
Intrinsic safety requirements8—30 VDCpower supply
Range:0-10ppm,0-100ppm,0-1000ppm
Resolution: 0.01ppm, 0.1ppm, 1ppm
Sampling method:
Diffuse type (if necessary, an external electric pump with a connector can be used)
Display mode:
7Section display,4Digit numberLCD,4colorLED
Response time:
sensorT90< 5Second, detector—1Minutes (standard)
accuracy:
0—100ppm> 98%
Calibration:
Magnetic suction, calibrated with a package100ppmIsobutene calibration
Operating Temperature
-20℃— 50℃(-4℉—122℉)
Working humidity:
0 — 95%RH(No condensation)
Weight and volume:
weight—Instrument:1.3kg(2.9lb)
Packing:1.47kg(3.2lb)
volume120×160×70mm(4.7″×6.3″×2.8″)
Note:
The above technical specifications are based on20℃,90%RHAnd the concentration is100ppmCalibration results under isobutene conditions.