Welcome Customer !

Membership

Help

Shanghai Zhongrong Yidian Co., Ltd
Custom manufacturer

Main Products:

instrumentb2b>Products

Three phase AC/DC instrument Transmitter calibration device (three in one)

NegotiableUpdate on 03/30
Model
Nature of the Manufacturer
Producers
Product Category
Place of Origin
Overview
Three phase AC/DC instrument Transmitter calibration device (three in one)
Product Details



Main indicators

>AC voltage 0-600V>AC current 0-20A
>DC voltage 0-1000V>DC current 0-20A
>Output capacity 20VA>Phase 0-360 degrees
>Frequency 45-65Hz>accuracy level 0.05/0.1

Main functions
1. 6.5-inch color LCD screen with clear display of chart characters;
2. It can semi automatically or manually inspect the basic errors of various power frequency meters (voltmeter, ammeter, power meter, frequency meter, power factor meter, phase meter) in the power system, as well as the changes caused by factors such as voltage, current, waveform, and power factor;
3. It can semi automatically or automatically verify the basic errors of electrical measurement transmitters (voltage transmitters, current transmitters, power transmitters, frequency transmitters, power factor transmitters, energy transmitters), as well as the changes caused by factors such as voltage, current, waveform, and power factor;
4. It can generate distortion waves of 2-31 harmonics, and the number, order, amplitude, and phase of harmonics on the fundamental wave can be programmed;
The working frequency band of the amplifier is 40Hz~1kHz, with good linearity. The current amplifier is a constant current source, and the voltage amplifier is a constant voltage source;
6. Equipped with an RS-232 interface, this device can be controlled by a computer through upper computer software for automatic or manual inspection, and the results can be processed and managed;
7. It can store 300 pieces of raw data from the tested instruments for reference and uploading;
8. The energy meter can be calibrated based on the load point or self selected point detected by the energy meter;

9. The device is lightweight and can be used in laboratories or carried for on-site verification;
10. This device has high precision, reliable working stability, and easy and flexible operation.
Technical indicators
1. AC voltage range: 50V, 100V, 200V, 400V, 600V, maximum output capacity: 20VA
2. AC current range
0.5A、1A 、2.5A、5A、10A、20A, Maximum output capacity 20VA
3. Range of AC voltage and current regulation0~130% FS, Adjust the fineness to 5 × 10-5
4. Accuracy of AC voltage and current at power frequency
0.05% FS
5. Active power (50V~400V) accuracy 0.05% FS
6. Reactive power (50V~400V) accuracy 0.1% FS
7. Phase accuracy of current to same phase voltage
0.05°
8. Frequency adjustment range45~65Hz, Adjust the fineness to 0.001Hz, with an accuracy of 5 × 10 for the set value-5
9. Phase adjustment range
0~359.99°Adjust the fineness by 0.01°
10. AC voltage and current output waveform distortion ≤ 0.3%
11. AC voltage, current, and power output stability ≤ 0.01% FS/60s
12. Harmonics 2-31, amplitude 0-20%, phase fineness of each harmonic 0.01
°N (where N is the harmonic order)
13. DC voltage range 75mV, 75V, 150V, 300V, 500V, 1000V, maximum output capacity 20W
14. DC current range 0.5A, 1A, 2.5A, 5A, 10A, 20A, maximum output capacity 20W
15. DC voltage and current regulation range 0-120% FS (excluding 1000V range), regulation fineness 5 × 10-5
16. DC voltage and current output stability ≤ 0.01% FS/60s (peak to peak)
17. Accuracy of DC voltage setting value 0.05% FS
18. Accuracy of DC current setting value 0.1% FS
19. DC 75mV output (load ≥ 5 Ω) accuracy 0.1% FS
20. DC measurement accuracy (transmitter output measurement) 0.02% FS (voltage 0~± 24V, current 0~± 24mA)
21. Comprehensive error of electric energy meter measurement (50V~400V) 0.05% (pulse output: three-phase rated value=6Hz)
22. Working power supply single-phase 220V ± 10%, 50Hz ± 5%;
23. Operating environment temperature 20
°C±10°CRelative humidity ≤ 85% RH
24. Volume weight 460 × 430 × 185mm, 20kg
Standard configuration details
1. Device body: 1 unit
2. One power cord
3. Test 1 set of wires and connectors
4. One dedicated DC mV testing line
5. One pulse test line
6. One data transmission cable
7. One copy of the user manual
8. One factory inspection report
9. One certificate of conformity
10. One aluminum alloy chassis