Thermtest TLS-100 Transient Line Heat Source InstrumentIt is a portable instrument used to measure the thermal conductivity and thermal resistance of various samples, including soil, rock, concrete, and polymers. Simply press the button to conduct the test and display the results immediately. the second generationTLS-100It has multiple upgrade functions, such as automatic recognition of sensors, automatic loading of test parameters, etc.

Transient line heat source(TLS)Instrument complianceASTM D5334andIEEE 442-2017Standard. The sensor needle is sealed in150mmThe100mmor50mmComposed of fine heating wires and temperature sensors in steel pipes, which can be inserted into the sample to be tested using a constant current source(q)Transfer heat to the sample and record the temperature rise during the defined time period. Thermal conductivity(k)Using the slope in the logarithmic plot of temperature rise and time(a)To calculate, the lower the slope, the higher the thermal conductivity of the sample; The higher the slope, the lower the thermal conductivity of the sample.

Thermtest TLS-100 Transient Line Heat Source InstrumentProduct Features
l Complies with international standards:ASTM D5334andIEEE 442-2017
l Portable, economical, and precise
l easy to use
l standard100mm/150mmSensors used for soft materials
l optional50mmSensors used for hard materials
TLS-100 specification
Method |
TLS-100 (Included) |
TLS-50 |
TLS-150 |
TLS-100 vCp |
|||
suitable material |
soil,paste,Powder and solid |
cement,Rocks and polymers |
soil,paste,Powder and solid |
soil,paste,Powder and solid |
|||
thermal conductivity(W/m·K) |
0.1 to 5 |
0.3 to 5 |
0.1 to 3 |
N/A |
|||
thermal resistance(mK/W) |
0.2 to 10 |
0.2 to 3.3 |
0.3 to 10 |
N/A |
|||
Volume specific heat capacity(MJ/m³K) |
N/A |
N/A |
N/A |
Up to 2.5 |
|||
Minimum sample size(mm) |
100 long, 50 diameter |
50 long, 50 diameter |
150 long, 50 diameter |
100 long, 50 diameter |
|||
Maximum sample size(mm) |
No restrictions |
No restrictions |
No restrictions |
No restrictions |
|||
Test time(minute) |
3 |
3 |
3 |
N/A |
|||
accuracy(thermal conductivity) |
5% |
5% |
5% |
15%* |
|||
reproducibility(thermal conductivity) |
2% |
2% |
2% |
2% |
|||
temperature range |
-40 to 100 |
||||||
Test standard |
ASTM D5334-22a, IEEE 442-1981 |
N/A |
ASTM D5334-14, IEEE 442-2017 |
N/A |
|||
*Specific heat